Pular para o conteúdo
Merck
Todas as fotos(1)

Key Documents

MFGB20SB3

Millipore

Hydrophobic Durapore® 0.22 µm, Millipak® Final Fill Capsule

Millipak® Final Fill 200, sterile; γ-irradiated

Faça loginpara ver os preços organizacionais e de contrato


About This Item

Código UNSPSC:
23151806

Materiais

polysulfone device
polysulfone support

Nível de qualidade

Agency

meets EP 2.6.14
meets JP 4.01
meets USP 85

esterilidade

irradiated
sterile; γ-irradiated

linha de produto

Millipak® Final Fill 200

Características

hydrophilic

fabricante/nome comercial

Millipak®

Parâmetros

10 psi max. differential pressure (0.7 bar) at 25 °C (Reverse)
200 L process volume
5.5 bar max. inlet pressure (80 psi) at 25 °C
7.61 L/min flow rate at 1.1 bar (ΔP)
80 psi max. differential pressure (5.5 bar) at 25 °C (Forward)

técnica(s)

sterile filtration: suitable

filter filtration area

1000 cm2

Impurezas

<0.25 EU/mL USP bacterial endotoxins (LAL test, per device; aqueous extraction)
<0.25 EU/mL bacterial endotoxins (LAL test, per device; aqueous extraction)

matriz

Durapore®

tamanho de poro

0.22 μm pore size

conector

(13 mm (1/2 in.) Hose Barb Inlet and Outlet)

Procurando produtos similares? Visita Guia de comparação de produtos

Categorias relacionadas

Descrição geral

Device Configuration: Gamma Gold Capsule

Embalagem

Double Easy-Open bag

Nota de preparo

Sterilization Method: Device integrity and retention was maintained after 3 autoclave cycles of 90 minutes at 126  °C. Devices can withstand a dose ≤40  kGy gamma exposure.

This product was manufactured with a Durapore® membrane which meets the criteria for a "non-fiber releasing" filter as defined in 21 CFR 210.3 (b)(6), validated based on large volume parenteral specifications as detailed in USP 788 Particulate Matter in Injections

Nota de análise

Bacterial Grow-Through: Quantitative retention of 1 x 107 CFU/cm2 Brevundimonas diminuta ATCC® 19146 per ASTM® F838 methodology.

Outras notas

Directions for Use
  • Organism Retention: Microorganism
  • Mode of Action: Filtration (size exclusion)
  • Application: BioProcessing
  • Intended Use: Reduction or removal of microorganism/bioburden
  • Instructions for Use: Please refer general guidelines section of user guide shipped with this product
  • Storage Statement: Please refer user guide shipped with this product
  • Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations

These products are manufactured in a facility which is certified to ISO 9001:2015 Quality Management Systems.

Informações legais

ASTM is a registered trademark of American Society for Testing and Materials
ATCC is a registered trademark of American Type Culture Collection
Durapore is a registered trademark of Merck KGaA, Darmstadt, Germany
MILLIPAK is a registered trademark of Merck KGaA, Darmstadt, Germany

Não está encontrando o produto certo?  

Experimente o nosso Ferramenta de seleção de produtos.


Certificados de análise (COA)

Busque Certificados de análise (COA) digitando o Número do Lote do produto. Os números de lote e remessa podem ser encontrados no rótulo de um produto após a palavra “Lot” ou “Batch”.

Já possui este produto?

Encontre a documentação dos produtos que você adquiriu recentemente na biblioteca de documentos.

Visite a Biblioteca de Documentos

Nossa equipe de cientistas tem experiência em todas as áreas de pesquisa, incluindo Life Sciences, ciência de materiais, síntese química, cromatografia, química analítica e muitas outras.

Entre em contato com a assistência técnica