HPLC Analysis of Benzodiazepines on Ascentis® Express Phenyl-Hexyl with Acetonitrile Mobile Phase
Materialien
Analytische Säule
Produkt-Nr.
Beschreibung
Preisangaben
Ascentis® Express Phenyl-Hexyl, 2,7 μm HPLC-Säule
2.7 μm particle size, L × I.D. 5 cm × 4.6 mmStandard
Produkt-Nr.
Beschreibung
Preisangaben
Alprazolam -Lösung
1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®Clonazepam -Lösung
1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®Diazepam -Lösung
1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®CONDITIONS
column
Ascentis Express Phenyl-Hexyl, 5 cm x 4.6 mm I.D., 2.7 μm particles (53348-U)
mobile phase
[A] water; [B] acetonitrile; (62:38, A:B)
flow rate
1.4 mL/min
pressure
2600 psi (179 bar)
column temp.
35 °C
detector
UV, 250 nm
injection
2 μL
sample
0.1 g/mL of each in 20% methanol
Beschreibung
Hinweis zur Analyse
The Phenyl-Hexyl phase of the Ascentis Express product line displays the best selectivity of this sample; selectivity is markedly affected by choice of organic modifier. The application also demonstrates feasibility of fast LC with Ascentis Express by virtue of low backpressure and high efficiency.
Rechtliche Hinweise