G005922
HPLC Analysis of Aflatoxins B1, B2, G1 and G2 in Corn on Ascentis® Express C18 after SPE using Supel™ Tox AflaZea and Fluorescence Detection
application for HPLC
About This Item
technique(s)
HPLC: suitable
test parameters
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: 25 g corn meal in 100 mL acetonitrile:water (84:16)
SPE tube/cartridge: Supel™ Tox AflaZea, 6 mL (55314-U)
sample addition: 2 mL
elution: collect sample in 5 mL silanized glass vial
eluate post-treatment: transfer 200 μL sample, add 300 μL TFA derivatization solution (70:20:10 water:TFA:acetic acid), heat in microreactor 20 min at 65 °C, cool, add 580 μL deionized water, vortex
column: Ascentis Express C18, 15 cm x 2.1 mm I.D., 2.7 μm particles (53825-U)
mobile phase: [A] water; [B] acetonitrile; (80:20, A:B)
flow rate: 0.350 mL/min
pressure: 325 psi (22.4 bar)
column temp.: 35 °C
detector: fluorescence, ex 360 nm, em 440 nm
injection: 20 μL
suitability
application for HPLC
application(s)
food and beverages
Analysis Note
Legal Information
Analito
Analytical column
Related product
Elija entre una de las versiones más recientes:
Certificados de análisis (COA)
Lo sentimos, en este momento no disponemos de COAs para este producto en línea.
Si necesita más asistencia, póngase en contacto con Atención al cliente
¿Ya tiene este producto?
Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.
Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.
Póngase en contacto con el Servicio técnico