G005880
GC Analysis of Polynuclear Aromatic Hydrocarbons (PAHs) in Grilled Hamburger on SPB®-608 (20 m x 0.18 mm I.D., 0.18 μm) after QuEChERS Cleanup using Supel™ QuE Z-Sep, Fast GC Analysis
suitable for GC, application for SPE
About This Item
technique(s)
gas chromatography (GC): suitable
solid phase extraction (SPE): suitable
test parameters
sample preparation: Dispersive (QuEChERS)
sample/matrix: grilled hamburger spiked with PAHs at 100 ng/g
extraction tube: Sodium chloride/magnesium sulfate (55294-U)
extraction process: add 25 mL acetonitrile, contents of unbuffered salt tube (55294-U); shake for 5 minutes; centrifuge for 5 minutes at 3200 rpm
clean-up tube: Supel QuE Z-Sep tube (55403-U)
clean-up process: transfer 3 mL of acetonitrile layer into a Supel QuE Z-Sep cleanup tube (55403-U); shake 1 minute; centrifuge for 3 minutes at 3400 rpm; remove 1 mL supernatant for GC-MS analysis
column: SPB-608, 20 m x 0.18 mm I.D. x 0.18 μm
oven: 60 °C (1 min), 25 °C/min to 275 °C, 10 °C/min to 300 °C (13 min)
inj. temp.: 265 °C
detector: MS-SIM
MSD interface: 300 °C
carrier gas: helium, 1.5 mL/min constant flow
injection: 1 μL splitless (1 min.)
liner: 4 mm I.D. FocusLiner™ with taper
sample: grilled hamburger extract
suitability
application for SPE
application(s)
food and beverages
Analysis Note
Other Notes
Legal Information
Choose from one of the most recent versions:
Certificados de análisis (COA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact Atención al cliente
¿Ya tiene este producto?
Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.
Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.
Póngase en contacto con el Servicio técnico