G005530
HPLC Analysis of Bisphenol A in Water on Ascentis® Express C18 after SPE using Supelclean™ ENVI-18
application for HPLC
About This Item
technique(s)
HPLC: suitable
test parameters
SPE tube/cartridge: Supelclean ENVI-18, 500 mg/6 mL glass SPE tube (54331-U)
condition: 1 mL 1% formic acid in acetonitrile, then 1 mL water
sample: 5 mL spiked water blank, concentration of 0.2 μg/mL
elution: 2 mL 1% formic acid in acetonitrile
eluate post-treatment: 1 mL aliquot of eluent, and evaporate/reconstitute to 0.5 mL in acetonitrile
column: Ascentis Express C18, 10 cm x 2.1 mm I.D., 2.7 μm particles (53823-U)
mobile phase: [A] water; [B] acetonitrile; (60:40, A:B)
flow rate: 0.4 mL/min
pressure: 3270 psi (256 bar)
column temp.: 35 °C
detector: UV, 230 nm
injection: 1 μL
suitability
application for HPLC
application(s)
environmental
Analysis Note
Legal Information
Analito
Analytical column
SPE tube or plate
Standard
Elija entre una de las versiones más recientes:
Certificados de análisis (COA)
Lo sentimos, en este momento no disponemos de COAs para este producto en línea.
Si necesita más asistencia, póngase en contacto con Atención al cliente
¿Ya tiene este producto?
Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.
Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.
Póngase en contacto con el Servicio técnico