749-0798
Semivolatiles
suitable for GC, application for SPE
About This Item
Productos recomendados
technique(s)
gas chromatography (GC): suitable
solid phase extraction (SPE): suitable
test parameters
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: acidified drinking water
SPE tube/cartridge: ENVI-18 DSK SPE Disk, 47 mm (57171)
condition: 5 mL dichloromethane; 5 mL methanol; 5 mL DI water
sample addition: 5 mL water sample
elution: 2 x 5 mL acetonitrile rinsing internal chamber of sample reservoir thoroughly
column: 5% phenyl polysiloxane, 30 m × 0.25 mm I.D., 0.25 μm (Supelco equivalent, SLB-5ms, 28471-U)
oven: 40 °C to 160 °C (3 min), then to 300 °C (3 min) at 6 °C/min
detector: MS, scan range m/z = 45-450
carrier gas: helium, 33 cm/sec
injection: 1 μL split/splitless, 1 min delay
suitability
application for SPE
application(s)
environmental
food and beverages
Application
Adjust 1 L drinking water to pH <2 with 6 N HCl. Add 5 mL methanol and mix thoroughly
Legal Information
Analito
Analytical column
Related product
Elija entre una de las versiones más recientes:
Certificados de análisis (COA)
Lo sentimos, en este momento no disponemos de COAs para este producto en línea.
Si necesita más asistencia, póngase en contacto con Atención al cliente
¿Ya tiene este producto?
Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.
Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.
Póngase en contacto con el Servicio técnico