技術
HPLC: suitable
測試參數
column: Hibar Purospher® STAR RP-18 endcapped, 150 x 4.6 mm, 3 μm (1.50470.0001)
mobile phase: [A] 30 mM ammonium phosphate, pH 5.3 with ammonium hydroxide in water; [B] Methanol; 97.5 [A] / 2.5 [B]
flow rate: 1.0 mL/min
pressure: 3060 psi 211 Bar
column temp.: 25 °C
detector: UV 254
injection: 5 μL
sample: 5 μg/mL of 12 nucleosides in 98:2 30mM ammonium phosphate pH 5.3 with ammonium hydroxide in water: methanol
適合性
application for HPLC
應用
life science and biopharma
分析報告
Nucleosides are structural subunits of nucleic acids, the macromolecules that convey genetic information in living cells. They contain a molecule of sugar (pentose) bonded to a nitrogen-containing organic ring compound. The nitrogen-containing compound is either a pyrimidine (cytosine, thymine, or uracil) or a purine (adenine or guanine). Nucleosides, derivatives and / or analogues can be used in many different industries and applications as in agrochemistry (herbicides, fungicides and insecticides), therapeutic drugs including a range of antiviral (HIV, hepatitis C virus and herpes simplex) and even antibiotics.
法律資訊
Purospher is a registered trademark of Merck KGaA, Darmstadt, Germany
分析物
描述
Pseudouridine
Cytidine
3-methylcytidine methosulfate
1-methyladenosine
Uridine
2-thiocytidine
5-methylcytidine
7-methylguanosine
2′-O-methylcytidine
Inosine
Guanosine
Ribothymidine
查看全部 (12)
標準
產品號碼
描述
訂價
我們的科學家團隊在所有研究領域都有豐富的經驗,包括生命科學、材料科學、化學合成、色譜、分析等.