UHPLC Analysis of Nucleosides on Purospher® STAR RP-18 endcapped

application for HPLC

登入查看組織和合約定價


About This Item

技術

HPLC: suitable

測試參數

column: Hibar Purospher® STAR RP-18 endcapped, 150 x 4.6 mm, 3 μm (1.50470.0001)
mobile phase: [A] 30 mM ammonium phosphate, pH 5.3 with ammonium hydroxide in water; [B] Methanol; 97.5 [A] / 2.5 [B]
flow rate: 1.0 mL/min
pressure: 3060 psi 211 Bar
column temp.: 25 °C
detector: UV 254
injection: 5 μL
sample: 5 μg/mL of 12 nucleosides in 98:2 30mM ammonium phosphate pH 5.3 with ammonium hydroxide in water: methanol

適合性

application for HPLC

應用

life science and biopharma

分析報告

Nucleosides are structural subunits of nucleic acids, the macromolecules that convey genetic information in living cells. They contain a molecule of sugar (pentose) bonded to a nitrogen-containing organic ring compound. The nitrogen-containing compound is either a pyrimidine (cytosine, thymine, or uracil) or a purine (adenine or guanine). Nucleosides, derivatives and / or analogues can be used in many different industries and applications as in agrochemistry (herbicides, fungicides and insecticides), therapeutic drugs including a range of antiviral (HIV, hepatitis C virus and herpes simplex) and even antibiotics.

法律資訊

Purospher is a registered trademark of Merck KGaA, Darmstadt, Germany

分析物

描述

    Pseudouridine
    Cytidine
    3-methylcytidine methosulfate
    1-methyladenosine
    Uridine
    2-thiocytidine
    5-methylcytidine
    7-methylguanosine
    2′-O-methylcytidine
    Inosine
    Guanosine
    Ribothymidine