LC/MS/MS Analysis of DL-d-Hydroxyglutaric Acid Enantiomers on Astec® CHIROBIOTIC® R

application for HPLC

登入查看組織和合約定價


About This Item

技術

HPLC: suitable

測試參數

column: CHIROBIOTIC R, 25 cm x 4.6 mm I.D., 5 μm particles (13024AST)
mobile phase: (A) ethanol:methanol (3:1, v:v); (B) 0.1% v/v TEA in water, adjusted to pH 4.5 with acetic acid; 75:25 (A:B)
flow rate: 0.4 mL/min
column temp.: 20 °C
detector: MS, ESI(-), 100-1000 m/z
injection: 5 μL
sample: each enantiomer, 100 μg/mL in water

適合性

application for HPLC

應用

clinical
forensics and toxicology
life science and biopharma

分析報告

Chiral 2-hydroxyglutarates (2-OHG) are important molecular signatures of both healthy and diseased biological cells, their specific biochemical pathways and inborn errors of metabolism. The chiral differentiation and quantification of D-2-OHG (R(-)-2-OHG) and L-2-OHG (S(+)-2-OHG) is key for characterizing neurometabolic disorders like the 2-hydroxyglutaric acidurias that cause neurological impairment early in life. The ability to distinguish and quantify the enantiomers of 2-OHG is therefore important to be able to track its disposition throughout the metabolic pathway of normal vs. mutated cells. We show here a rapid and sensitive method to separate and measure D-2-OHG and L-2-OHG using LC-MS/MS detection. An Astec® CHIROBIOTIC R (ristocetin chiral selector) column was run under polar ionic mobile phase conditions. Using tandem MS eliminated two fundamental analytical challenges with these compounds: the lack of a UV-absorbing chromophore in the 2-OHG molecule, and the abundance of competing low molecular weight acids in urine. The upper panel in the chromatogram shows the separation of D-2-OHG and L-2-OHG obtained during method development. The lower panel shows the pure L-2-hydroxyglutaric acid.
Besides providing the necessary enantioselectivity, Astec® CHIROBIOTIC columns have the advantage of operating in aqueous and polar organic mobile phases that are amenable to polar drugs and metabolites. Their ionic interactions promote analyte ionization which enhances sensitivity in ESI detection. In addition to the CHIROBIOTIC columns, Sigma-Aldrich also provides the high purity mobile phase agents, racemic and pure chiral 2-OHG enantiomers and other compounds in the isocitrate dehydrogenase (IDH) pathway.

法律資訊

Astec is a registered trademark of Merck KGaA, Darmstadt, Germany
CHIROBIOTIC is a registered trademark of Sigma-Aldrich Co. LLC

分析物

描述

    L-α-Hydroxyglutaric acid disodium salt, ≥98.0% (GC)
    D-α-Hydroxyglutaric acid disodium salt, ≥95% (GC)

從最近期的版本中選擇一個:

分析證明 (COA)

Lot/Batch Number

抱歉,我們目前沒有在線上提供本產品的COA

如果您需要協助,請聯絡 客戶支援

已經擁有該產品?

您可以在文件庫中找到最近購買的產品相關文件。

存取文件庫

我們的科學家團隊在所有研究領域都有豐富的經驗,包括生命科學、材料科學、化學合成、色譜、分析等.

聯絡技術服務