1.04140
Singlepath® Salmonella
suitable for lateral flow assay, input: food(s), for Salmonella spp.
Synonym(s):
Rapid Salmonella Food Test
Sign Into View Organizational & Contract Pricing
All Photos(4)
About This Item
Agency
according to AOAC® PTM
Quality Level
product line
Singlepath®
shelf life
15 mo.
packaging
pack of 25 strips
technique(s)
lateral flow assay: suitable
microbe id | immunological: suitable
input
food(s)
application(s)
food and beverages
pathogen testing
storage temp.
2-8°C
suitability
Citrobacter spp.
Salmonella spp.
General description
Singlepath® Salmonella is a rapid diagnostic test, designed to detect Salmonella bacteria in various food samples. Salmonella is a genus of bacteria that can cause human infections, leading to symptoms such as diarrhea, fever, and abdominal cramps.The Singlepath® Salmonella is an immunochromatographic rapid test based on gold-labeled antibodies to detect the presence of a specific antigen, in this case, the Salmonella bacteria. The test includes a nitrocellulose membrane strip that is coated with antibodies specific to Salmonella antigens. When the sample is applied to the test strip, any Salmonella antigens present will bind to the antibodies on the strip, resulting in the development of a visible test line. The test takes approximately 20 minutes to complete, and the results are interpreted based on the presence or absence of the test line. The detection limit can be regarded as less than one colony-forming unit in a 25 g food sample, depending on the serogroup. No additional equipment expenditure is needed.
Application
SINGLEPATH® Salmonella is an immunochromatographic rapid test intended for use in microbiology laboratories analyzing food and animal feeds for the presumptive qualitative detection of Salmonella spp. from food matrices such as meats (raw ground beef and raw ground turkey), spices (black pepper), dairy (dried skimmed milk), dried foods (coconut) and seafood (cooked, peeled frozen prawns).
Features and Benefits
- Rapid and reliable method
- High sensitivity and specificity compared to conventional culture methods
- Can be performed without the need for specialized equipment or personnel.
Linkage
Replaces: 1.04140.0002
Analysis Note
Performance test (Salmonella enteritidis NCTC 5188 (WDCM 00119)): +
Performance test (Salmonella typhimurium ATCC 14028 (WDCM 00031)): +
Performance test (Escherichia coli ATCC 25922 (WDCM 00013)): -
Performance test (Citrobacter freundii ATCC 8090): -
Performance test (Salmonella typhimurium ATCC 14028 (WDCM 00031)): +
Performance test (Escherichia coli ATCC 25922 (WDCM 00013)): -
Performance test (Citrobacter freundii ATCC 8090): -
Legal Information
AOAC is a registered trademark of AOAC International
SINGLEPATH is a registered trademark of Merck KGaA, Darmstadt, Germany
Storage Class Code
11 - Combustible Solids
Certificates of Analysis (COA)
Search for Certificates of Analysis (COA) by entering the products Lot/Batch Number. Lot and Batch Numbers can be found on a product’s label following the words ‘Lot’ or ‘Batch’.
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
Our team of scientists has experience in all areas of research including Life Science, Material Science, Chemical Synthesis, Chromatography, Analytical and many others.
Contact Technical Service