GC Analysis of PCBs in Transformer Oil on SPB®-5 after SPE using Supelclean LC-Florisil® and LC-Si

suitable for GC, application for SPE

로그인조직 및 계약 가격 보기


About This Item

기술

gas chromatography (GC): suitable
solid phase extraction (SPE): suitable

검사 파라미터

sample preparation: SPE (Solid Phase Extraction)
sample/matrix: used transormer oil containing 50 ppb Aroclor 1254.
SPE tube/cartridge: Supelclean LC-Florisil, 1.0 g/6 mL (57057)
SPE tube/cartridge: Supelclean LC-Si, 1.0 g/6 mL (57051)
condition: 0.5 mL isooctane (Florisil tube)
sample addition: 0.2 g transformer oil or 0.1 g waste oil
washing: 5 x 2 mL isooctane
column: SPB-5, 30 m × 0.32 mm I.D., 0.25 μm film (24048)
oven: 40 °C (4 min) to 300 °C at 10 °C/min, hold 5 min
detector: ECD
carrier gas: nitrogen
injection: 1 μL, splitless (30 sec delay), then split (50:1)

적합성

application for SPE

응용 분야

environmental

애플리케이션

Connect Supelclean LC-Si SPE tube (1 g/6 mL) in series to male luer outlet of Supelclean LC-Florisil SPE tube (1 g/6 mL) using an SPE Tube Adapter (57020-U). Apply up to 0.2 g transformer oil or 0.1 g waste oil to upper frit of Florisil SPE tube. Pass 5 x 2 mL isooctane through both Florisil and LC-Si SPE tubes. Oil interferences will be retained on SPE tubes whereas PCBs will elute with isooctane fractions. Collect and combine isooctane fractions using 10 mL volumetric flask. Analyze isooctane extract using GC-ECD.

법적 정보

Florisil is a registered trademark of U.S. Silica Company
SPB is a registered trademark of Merck KGaA, Darmstadt, Germany
Supelclean is a trademark of Sigma-Aldrich Co. LLC

가장 최신 버전 중 하나를 선택하세요:

시험 성적서(COA)

Lot/Batch Number

죄송합니다. 지금은 이 제품에 대한 COA이(가) 온라인에서 제공되지 않습니다.

도움이 필요하시면 연락하세요. 고객 지원 부서

이 제품을 이미 가지고 계십니까?

문서 라이브러리에서 최근에 구매한 제품에 대한 문서를 찾아보세요.

문서 라이브러리 방문

자사의 과학자팀은 생명 과학, 재료 과학, 화학 합성, 크로마토그래피, 분석 및 기타 많은 영역을 포함한 모든 과학 분야에 경험이 있습니다..

고객지원팀으로 연락바랍니다.