In order to assess any variations in LC-MS-based metabolomics analysis data, it is recommended to add an additional QC (quality control) sample at the beginning of every analytical experiment batch. It is recommended that QC be injected every 4-10 injections into the workflow steps. The QC sample evaluates any drifting phenomenon like intensity values, ion suppression, or any changes in the retention time of the material’s peaks in the LC-MS data. The Polar Metabolites QC Mix is a ready to use solution for mass spectroscopy workflows. By utilizing the QC mix, the user will be able to assess drifting and ion suppression phenomenon. The Polar Metabolites QC Mix is comprised of 8 components (Uridine, Adenosine, Nicotinic acid, Biotin, Phenylacetylglutamine, L-phenylalanine, Cytidine and L-Tryptophan) which covers ~ 100-500 Da mass range.
특징 및 장점
- The mix was designed for ESI (+/-) detection.
- The mixture comes as a ready to use solution(~9:1 ACN:10mM ammonium formate, inwater) format in crimp top (Silicone/PTFE liner) amber vial, which is suitable for use with most of the LC-MS autosamplers.
- The recommended storage temperature of the mixture is 2-8 °C, thawing of the sample is not required.
적합성
The Polar Metabolites QC Mix is suitable for mass spectroscopy
신호어
Danger
유해 및 위험 성명서
Hazard Classifications
Acute Tox. 4 Dermal - Acute Tox. 4 Inhalation - Acute Tox. 4 Oral - Eye Irrit. 2 - Flam. Liq. 2
Storage Class Code
3 - Flammable liquids
WGK
WGK 3
Flash Point (°F)
35.6 °F - closed cup
Flash Point (°C)
2.0 °C - closed cup