추천 제품
Agency
EP
USP
Quality Level
형태
liquid
특징
ready-to-use
포장
ampule of 2 mL
응용 분야
microbiology
저장 온도
2-8°C
적합성
Bacillus spp.
일반 설명
The Sterikon® plus Bioindicator contains a nutrient broth, sugar, a pH indicator, and spores of a non-pathogenic organism, Geobacillus stearothermophilus ATCC 7953 (sporulation optimized) in a single ampoule. The spores are completely killed upon heating in compressed steam at a temperature of 121 ± 0.5°C (245 kPa), after 15 minutes. At lower temperatures or lower exposure times, a small number of spores can survive at and can grow upon incubation. Sterikon® ampoules are placed into the autoclave along with the batch to be autoclaved. The efficiency of the sterilization process is checked by the incubation of the ampoules. Red (or red-violet) color of the ampoule contents indicates no growth of Geobacillus stearothermophilus and adequate sterilization, whereas the yellow (or yellow-orange) color of the ampoule contents indicates growth and inadequate sterilization.
애플리케이션
Sterikon® plus Bioindicator System is used to monitor the effectiveness of the autoclaving by steam sterilization for 15 minutes at 121°C. It is suitable for checking the sterilization efficiency of the culture media products used for testing of pharmaceuticals (especially drugs in ampoule form), food and water samples, along with canned foods and other products to be autoclaved.
결합
Replaces: 1.10274.0003
분석 메모
Appearance: opalescent and reddish violet
Number of viable spores per ampoule: 5.0E+05 - 1.0E+07
Growth (35 °C): no growth
Growth (60 °C): growth
Performance test (6 min, 121 °C): Survival time 6 min/121 °C
Performance test (15 min, 121 °C): Kill time 15 min/121 °C
D value (121 °C): 1.5 - 2.0 min
Z value: 7.0 - 10.0 °C
Identity: Geobacillus stearo-thermophilus ATCC 7953
The determination of the D-value is based on the Stumbo-Murphy-Cochran method.
The identity control is carried out via MicroSeq-System.
Organism: Geobacillus stearothermophilus ATCC 7953 (sporulation optimized).
Ampoules: 2 ml.
Storage temperature: +2° to +8°C.
Number of viable spores per ampoule: 5.0E+05 - 1.0E+07
Growth (35 °C): no growth
Growth (60 °C): growth
Performance test (6 min, 121 °C): Survival time 6 min/121 °C
Performance test (15 min, 121 °C): Kill time 15 min/121 °C
D value (121 °C): 1.5 - 2.0 min
Z value: 7.0 - 10.0 °C
Identity: Geobacillus stearo-thermophilus ATCC 7953
The determination of the D-value is based on the Stumbo-Murphy-Cochran method.
The identity control is carried out via MicroSeq-System.
Organism: Geobacillus stearothermophilus ATCC 7953 (sporulation optimized).
Ampoules: 2 ml.
Storage temperature: +2° to +8°C.
법적 정보
STERIKON is a registered trademark of Merck KGaA, Darmstadt, Germany
Storage Class Code
12 - Non Combustible Liquids
WGK
WGK 1
Flash Point (°F)
Not applicable
Flash Point (°C)
Not applicable
시험 성적서(COA)
제품의 로트/배치 번호를 입력하여 시험 성적서(COA)을 검색하십시오. 로트 및 배치 번호는 제품 라벨에 있는 ‘로트’ 또는 ‘배치’라는 용어 뒤에서 찾을 수 있습니다.
자사의 과학자팀은 생명 과학, 재료 과학, 화학 합성, 크로마토그래피, 분석 및 기타 많은 영역을 포함한 모든 과학 분야에 경험이 있습니다..