コンテンツへスキップ
Merck
  • Measurement of aflatoxin and aflatoxin metabolites in urine by liquid chromatography-tandem mass spectrometry.

Measurement of aflatoxin and aflatoxin metabolites in urine by liquid chromatography-tandem mass spectrometry.

Journal of analytical toxicology (2007-06-21)
Robert A Everley, Frederic L Ciner, Dongliang Zhan, Peter F Scholl, John D Groopman, Timothy R Croley
要旨

Automated immunoaffinity solid-phase extraction followed by liquid chromatography-tandem mass spectrometry and chemical analogue internal standardization is employed to detect and quantify the aflatoxins AFB(1), AFB(2), AFG(1), AFG(2), and the metabolites AFM(1) and AFP(1) in urine. The dynamic range of the method is nearly three orders of magnitude with limits of detection in the low femtogram on column range. The method was validated over a 12-day period by eight analysts. This method is suitable for agricultural, forensic, and public health laboratories during an accidental outbreak or a chemical terrorism event where a rapid and accurate diagnosis of aflatoxicosis is needed.

材料
製品番号
ブランド
製品内容

Sigma-Aldrich
アフラトキシンG2
Supelco
アフラトキシンG2 溶液, 0.5 μg/mL in acetonitrile, analytical standard
アフラトキシンG2 溶液, 3.80 μg/g in acetonitrile, ERM®, certified reference material
Supelco
アフラトキシンG2