Piezoelectric-excited millimeter-sized cantilever biosensors.

Methods in molecular biology (Clifton, N.J.) (2009-01-23)
Raj Mutharasan
要旨

In this chapter, method of fabricating a cantilever biosensors and their use in measuring the presence of a protein is described. There are many variations in construction of a cantilever sensor. A simple and an easy version is described in this chapter. The specificity of the sensor is obtained by immobilizing an antibody specific to the antigen of interest. The piezoelectric cantilever sensors are very sensitive and can easily detect a 60 kDa protein at 100 fg/mL concentration. Experimental procedure for carrying out detection of a target analyte is outlined and a sample set of results is included.

材料
製品番号
ブランド
製品内容

Sigma-Aldrich
プロテインG′, recombinant, expressed in E. coli, lyophilized powder
Sigma-Aldrich
抗ウシアルブミン (BSA)−アガロース ウサギ宿主抗体, IgG fraction of antiserum, PBS suspension