HPLC Analysis of Acetaminophen, Caffeine, Chlorpheniramine, Ibuprofen, Naproxen, Phenylephrine, and Pseudoephedrine on Ascentis® Express C18

application for HPLC

Autenticatiper visualizzare i prezzi riservati alla tua organizzazione & contrattuali


About This Item

tecniche

HPLC: suitable

Parametri analitici

column: Ascentis Express C18, 5 cm x 2.1 mm I.D., 2.7 μm particles (53822-U)
mobile phase: [A] 5 mM ammonium phosphate monobasic, pH 2.0 with phosphoric acid: acetonitrile (98:2); [B] acetonitrile;
gradient: 0 to 68% B in 3 min; held at 68% B for 2 min
flow rate: 0.4 mL/min
column temp.: 35 °C
detector: UV, 210 nm
injection: 1 μL
sample: 100 μg/mL in 95:5, water: methanol

Compatibilità

application for HPLC

applicazioni

pharmaceutical (small molecule)

Risultati analitici

This application demonstrates the suitability of Ascentis Express C18 for the efficient separation of various over the counter drugs.

Note legali

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

Analita

Descrizione

    Maleic acid
    Acetaminophen
    Caffeine
    Naproxen
    Ibuprofen

Scegli una delle versioni più recenti:

Certificati d'analisi (COA)

Lot/Batch Number

It looks like we've run into a problem, but you can still download Certificates of Analysis from our Documenti section.

Se ti serve aiuto, non esitare a contattarci Servizio Clienti

Possiedi già questo prodotto?

I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.

Visita l’Archivio dei documenti

Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..

Contatta l'Assistenza Tecnica.