G001056
HPLC Analysis of Barbiturates in Serum on Discovery® C18 after SPE using Discovery® DSC-18Lt
application for SPE, application for HPLC
About This Item
tecniche
HPLC: suitable
solid phase extraction (SPE): suitable
Parametri analitici
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: 0.5 mL porcine serum spiked with 0.5 μg/mL or 1.0 μg/mL each analyte then diluted with 0.5 mL water
SPE tube/cartridge: Discovery DSC-18Lt, 500 mg/3 mL (52613-U)
condition: 2 mL methanol; 2 mL DI water
sample addition: 1 mL at 0.75 mL/min
washing: 2 mL 5% methanol, then vacuum or air dry for 5-10 min
elution: 1-2 mL methanol
eluate post-treatment: dry eluate with nitrogen purge (40 °C; 15-20 min), reconstitute in 20 μL mobile phase
column: Discovery C18, 15 cm × 4.6 mm, 5 μm preceded by a 2 cm C18 guard column and 0.5 μm frit filter (504955)
mobile phase: [A] methanol; [B] water (40:60, A:B)
flow rate: 1 mL/min
column temp.: 30 °C
detector: UV, 214 nm
injection: 30 μL, diluted porcine serum extract
Compatibilità
application for HPLC
application for SPE
applicazioni
forensics and toxicology
Risultati analitici
Note legali
Analita
Colonna analitica
Colonnina o piastra per SPE
Standard
Scegli una delle versioni più recenti:
Certificati d'analisi (COA)
Ci dispiace, ma al momento non ci sono COA disponibili online per questo prodotto.
Se ti serve aiuto, non esitare a contattarci Servizio Clienti
Possiedi già questo prodotto?
I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.
Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..
Contatta l'Assistenza Tecnica.