US EPA Method 610: GC Analysis of Polynuclear Aromatic Hydrocarbons (PAHs) on SLB®-5ms

suitable for GC

登录查看公司和协议定价


About This Item

技術

gas chromatography (GC): suitable

測試參數

column: SLB-5ms, 30 m x 0.25 mm I.D., 0.25 μm (24103-U)
oven: 70 °C (0.2 min), 25 °C/min to 265 °C, 5 °C/min to 315 °C (2 min)
inj. temp.: 250 °C
detector: FlD, 350 °C
carrier gas: helium, 1.3 mL/min, constant
injection: 2.0 μL pulsed splitless (30 psi until 0.2 min)
liner: 4 mm I.D. single taper
sample: 1 μg/mL PAH standard in methylene chloride

應用

environmental

分析報告

US EPA Method 610 describes the analysis of polynuclear aromatic hydrocarbons (commonly referred to as PAHs or PNAs) by both HPLC and GC. The method designates the use of a packed GC column, however capillary columns will provide better resolution, baseline stability, and longer lifetime. This application illustrates the separation of the target compound list in this method on the SLB-5ms capillary GC column. A pressure pulsed injection was used to improve response of the heavier PAHs, and to reduce .the larger vapor volume resulting from a 2 μL injection.

法律資訊

SLB is a registered trademark of Merck KGaA, Darmstadt, Germany

分析物

说明

    1-Fluoronaphthalene
    Naphthalene
    2-Fluorobiphenyl
    Acenaphthylene
    Acenaphthene
    Fluorene
    Phenanthrene
    Anthracene
    Fluoranthene
    Pyrene
    Chrysene
    Benzo[b]fluoranthene
    Benzo[k]fluoranthene
    Indeno[1,2,3-cd]pyrene
    Benzo[ghi]perylene
    查看所有结果 (15)