技術
gas chromatography (GC): suitable
測試參數
column: SUPELCOWAX 10, 60 m x 0.25 mm I.D., 0.25 μm (24081)
oven: 50 °C (3 min), 15 °C/min to 265 °C (5 min)
inj. temp.: 250 °C
detector: FID, 265 °C
carrier gas: helium, 30 cm/sec
injection: 0.5 μL, 200:1 split
liner: 4 mm I.D., splitless type, single taper wool packed FocusLiner™ design
sample: reformulated unleaded gasoline
應用
petroleum
分析報告
This application illustrates the analysis of a sample of reformulated gasoline on the SUPELCOWAX 10 column. Due to the polar nature of the phase, the aliphatics elute earlier than the aromatics. There is some overlap in the elution ranges, however it is not as severe as that exhibited by a nonpolar column.
法律資訊
FocusLiner is a trademark of SGE Analytical Science Pty Ltd.
分析物
说明
Ethanol denatured
Benzene
Decane
Toluene
Undecane
Ethylbenzene
p-Xylene
m-Xylene
Dodecane
o-Xylene
Tridecane
1,2,4-Trimethylbenzene
1,2,3-Trimethylbenzene
Tetradecane
Durene
Naphthalene
2-Methylnaphthalene
1-Methylnaphthalene solution