主要文件

G005521

GC Analysis of Phthalate Esters in Perfume on SLB®-5ms (20 m x 0.18 mm I.D., 0.18 μm), Fast GC Analysis

suitable for GC

登录查看公司和协议定价


About This Item

技術

gas chromatography (GC): suitable

測試參數

column: SLB-5ms, 20 m x 0.18 mm I.D., 0.18 μm (28564-U)
oven: 60 °C (1 min), 10 °C/min to 330 °C (10 min)
inj. temp.: 250 °C
carrier gas: helium, 0.6 mL/min, constant
liner: 4 mm ID focus liner with taper
sample: cologne sample spiked at 100 μg/mL with various phthalate esters

應用

cleaning products
cosmetics
personal care

一般說明

Diethyl phthalate was detected in the unspiked sample.

分析報告

Phthalate esters are commonly used as plasticizers and additives for various materials. In the case of some fragrances, diethyl phthalate is used as a solvent and deluent. Rising concern with the potential health effects of exposure to phthalate esters has lead to some controversy over their use in fragrances and other personal care products. This application illustrates the use of GC-MS to analyze a cologne sample that has been spiked at 100 ug/ml with a list of phthalate esters. The unspiked sample was found to contain diethyl phthalate.
The SLB-5ms column was chosen for the analysis due to its low bleed/MS compatibility. The nonpolar character of the phase resulted in elution of the lower boiling constituents in the cologne sample prior to most of the phthalate esters. A 20 m x 0.18 mm ID x 0.18 μm column was used to reduce overall analysis time.