推荐产品
ligand
native protein A
包裝
pack of 1 × 5 mL
製造商/商標名
Cytiva 17-0403-01
儲存條件
(20% Ethanol)
參數
flow rate
<20 mL/min flow rate
42 psi
床尺寸
16 mm × 25 mm
床體積
5 mL
柱內徑
16 mm
基質
highly cross-linked 6% agarose
樣品平均尺寸
34 μm
cleaning
2-10
工作範圍
3-9
儲存溫度
2-8°C
正在寻找类似产品? 访问 产品对比指南
一般說明
HiTrap® Protein A HP is a prepacked ready to use, column for preparative purification of monoclonal and polyclonal antibodies. The special design of the column, together with the medium provide fast, simple and easy separations in a convenient format.
特點和優勢
- High binding capacitiy of 20 mg IgG/mL medium and fast kinetics.
- 34 μm bead size for high-performance, high-resolution purifications.
- Designed for use with syringe, pump, or chromatography system such as AKTA design or FPLC™ System.
- Stable over pH 3 to 9 for use, and 2 to 10 for cleaning.
- Robust, easy-to-use method.
儲存和穩定性
分析報告
其他說明
法律資訊
訊號詞
Warning
危險聲明
儲存類別代碼
3 - Flammable liquids
其他客户在看
商品
This page shows how to purify IgM antibodies by affinity chromatography using HiTrap IgM Purification HP.
This page shows how to separate IgG antibodies by affinity chromatography using HiTrap® Protein G HP columns from GE Healthcare.
This page shows how to solve practical problems that may occur when running an affinity chromatography column.
实验方案
This page shows general instructions for affinity purification using HiTrap columns from GE Healthcare.
Protein A is derived from a strain of Staphylococcus aureus and contains five regions that bind to the Fc region of IgG. As an affinity ligand, protein A is coupled to Sepharose so that these regions are free to bind IgG.
This page shows how to convert between linear flow and volumetric flow rates in affinity chromatography.
This page shows how to perform column packing and preparation for affinity chromatography when using Tricorn™ or XK columns available from GE Healthcare.