G006299
LC/MS Analysis of Opioid Glucuronide Metabolites in Urine on Ascentis® Express F5 after Solid Phase Extraction (SPE) using Supel™-Select HLB
application for SPE, application for HPLC
About This Item
técnicas
HPLC: suitable
solid phase extraction (SPE): suitable
parámetros analíticos
sample preparation: Solid Phase Extraction
sample/matrix: urine spiked with drugs and metabolites at 300 ng/mL
SPE well plate: Supel-Select HLB 96-well plate, 60 mg/well (575662-U)
condition: 1 mL methanol followed by 1 mL water
sample addition: 1 mL hydrolyzed urine
washing: 1 mL water followed by 1 mL 25% methanol
elution: 1 mL acetonitrile
column: Ascentis Express F5, 10 cm x 2.1 mm I.D., 2.7 μm particles (53569-U)
mobile phase: [A] 5 mM ammonium formate, pH 4.2 (adjusted with formic acid); [B] 5 mM ammonium formate in (95:5 acetonitrile:water) pH 5.2 (adjusted with formic acid)
gradient: 15% B held for 1 min, to 70% B in 2 min, held at 70% B for 7 min
flow rate: 0.4 mL/min
column temp.: 35 °C
detector: ESI(+) TOF, full scan, m/z 100-1000
injection: 3 μL
sample: 300 ng/mL
idoneidad
application for HPLC
application for SPE
aplicaciones
clinical
forensics and toxicology
Nota de análisis
Información legal
Analito
Estándar
columna analítica
tubo o placa de SPE
Elija entre una de las versiones más recientes:
Certificados de análisis (COA)
Lo sentimos, en este momento no disponemos de COAs para este producto en línea.
Si necesita más asistencia, póngase en contacto con Atención al cliente
¿Ya tiene este producto?
Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.
Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.
Póngase en contacto con el Servicio técnico