Saltar al contenido
Merck
Todas las fotos(1)

Key Documents

G005524

US EPA Method 524.2: GC Analysis of Volatiles on SPB-624 after Purge & Trap using "K" Trap

suitable for GC

Iniciar sesiónpara Ver la Fijación de precios por contrato y de la organización


About This Item

técnicas

gas chromatography (GC): suitable

parámetros analíticos

sample/matrix: each analyte at 10 ppb in 25 mL water
purge trap: VOCARB 3000 "K" (24940-U)
dry purge: 2 min
desorption preheat: 205 °C
desorption process: 40 mL/min at 210 °C for 2 min
bake: 260 °C for 10 min
transfer line/valve temp.: 110 °C
sample preparation: purge is 40 mL/min at 25 °C for 11 min
column: SPB-624, 30 m x 0.25 mm I.D., 1.4 μm (24255)
oven: 40 °C (2 min), 7 °C/min to 135 °C, 30 °C/min to 230 °C (3 min)
carrier gas: helium, 1.2 mL/min
injection: 30:1 split
liner: 0.75 mm I.D. SPME

aplicaciones

environmental
food and beverages

Nota de análisis

Method 524.2 describes the analysis of volatile organic compounds in surface, ground, and drinking water. This application illustrates the analysis of the compounds listed in this method on a 30m x 0.25 mm I.D., 1.4 μm SPB-624 column using purge and trap / GC-MS.

Analito

Descripción

    m-Xylene
    p-Xylene

Elija entre una de las versiones más recientes:

Certificados de análisis (COA)

Lot/Batch Number

Lo sentimos, en este momento no disponemos de COAs para este producto en línea.

Si necesita más asistencia, póngase en contacto con Atención al cliente

¿Ya tiene este producto?

Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.

Visite la Librería de documentos

Nuestro equipo de científicos tiene experiencia en todas las áreas de investigación: Ciencias de la vida, Ciencia de los materiales, Síntesis química, Cromatografía, Analítica y muchas otras.

Póngase en contacto con el Servicio técnico