Passa al contenuto
Merck
Tutte le immagini(1)

Documenti

G004366

HPLC Analysis of Verapamil and Metabolites in Rat Plasma on Ascentis® Express C18 after Phospholipid Removal using HybridSPE®-Phospholipid

application for SPE, application for LC-MS

Autenticatiper visualizzare i prezzi riservati alla tua organizzazione & contrattuali


About This Item

tecniche

LC/MS: suitable
solid phase extraction (SPE): suitable

Parametri analitici

SPE well plate: HybridSPE-Precipitation 96-well Plate, 50 mg/well (575656-U)
sample preparation: SPE (Solid Phase Extraction)
column: Ascentis Express C18, 5 cm x 2.1 mm I.D., 2.7 μm particles (53822-U)
sample/matrix: Rat plasma spiked with verapamil and metabolites at 10 ng/mL
mobile phase: [A] water; [B] acetonitrile, pH 2.55 adjust with formic acid (30:70, A:B)
flow rate: 0.6 mL/min
column temp.: 35 °C
detector: TOF/MS
injection: 1 μL
sample addition: 100 μL spiked rat plasma followed by 300 μL 1% formic acid in acetonitrile
elution: apply vacuum

Compatibilità

application for LC-MS
application for SPE

applicazioni

clinical

Note legali

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany
HybridSPE is a registered trademark of Merck KGaA, Darmstadt, Germany

Choose from one of the most recent versions:

Certificati d'analisi (COA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact Servizio Clienti

Possiedi già questo prodotto?

I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.

Visita l’Archivio dei documenti

Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..

Contatta l'Assistenza Tecnica.