G000594
HPLC Analysis of Bronchodilators Theophylline and Other Caffeine Metabolites in Serum on Discovery® RP-AmideC16 after SPE using Discovery® DSC-18
application for SPE, application for HPLC
About This Item
tecniche
HPLC: suitable
solid phase extraction (SPE): suitable
Parametri analitici
sample preparation: SPE (Solid Phase Extraction)
SPE tube/cartridge: Discovery® DSC-18, 500 mg/3 mL (52603-U)
condition: 2 mL methanol, then 2 mL water
sample preparation: 1 mL porcine serum spiked with 0.1 μg/mL, 0.50 μg/mL, or 1.0 μg/mL each analyte
sample preparation: 1 mL methanol; evaporate to dryness with nitrogen stream at room temperature; reconstitute in 200 μL mobile phase containing 0.2 μg/mL or 7-methylxanthine (IS)
sample preparation: 2 mL 5% methanol in water; dry tube 10 min with nitrogen stream
column: Discovery RP-AmideC16, 15 cm × 4.6 mm, 5 μm particles, preceded by a 2 cm RP-AmideC16 guard column and 0.5 μm frit filter (505013)
mobile phase: [A] Methanol; [B] 1% acetic acid (17:38, A:B)
flow rate: 1 mL/min
column temp.: 30 °C
detector: UV, 272 nm
injection: 20 μL reconstituted porcine serum extract
Compatibilità
application for HPLC
application for SPE
applicazioni
clinical
pharmaceutical (small molecule)
Note legali
Analita
Colonna analitica
Colonnina o piastra per SPE
Scegli una delle versioni più recenti:
Certificati d'analisi (COA)
Ci dispiace, ma al momento non ci sono COA disponibili online per questo prodotto.
Se ti serve aiuto, non esitare a contattarci Servizio Clienti
Possiedi già questo prodotto?
I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.
Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..
Contatta l'Assistenza Tecnica.