Passa al contenuto
Merck
Tutte le immagini(2)

Key Documents

P2B030A01

Millipore

Pellicon® 2 Mini Modulo per ultrafiltrazione Biomax®

This product is being discontinued as notified in July 2022. Equivalent and new generation Pellicon® 3 cassettes are available for ordering.

Sinonimo/i:

Pellicon® 2 Mini Ultrafiltration Module Biomax-30 A 0.1 m2

Autenticatiper visualizzare i prezzi riservati alla tua organizzazione & contrattuali


About This Item

Codice UNSPSC:
23151806

Materiali

polyethersulfone
polypropylene screen
polyurethane adhesive

Livello qualitativo

Produttore/marchio commerciale

Pellicon®

Parametri

0.33 bar max. transmembrane pressure (5 psig) (reverse)
4-6 L/min recirculation rate at 1.4 bar (20 psi)
7 bar max. transmembrane pressure (100 psig) at 30 °C (forward)

tecniche

analytical sample preparation: suitable
protein purification: suitable

Altezza

1.5 cm (0.6 in.)

Lunghezza

21 cm (8.3 in.)

Larghezza

5.6 cm (2.2 in.)

Superficie filtrante

0.1 m2

membrane nominal area

0.1 m2

Estraibili (gravimetria)

≤1500 mg/m2

Matrice

Biomax®

Dimensione pori

30 kDa NMWCO

pH

1-14

Cerchi prodotti simili? Visita Guida al confronto tra prodotti

Descrizione generale

The Pellicon® 2 Ultrafiltration BIOMAX 30 Mini Cassette sets the standard in cassette-style tangential flow filtration devices. P2B030A01 was designed with process scale-up and applications development in mind.

Millipore′s BIOMAX 30 polyethersulfone membrane has void-free structures that guard against leakage of solutes through microdefects normally associated with voids beneath the thin skins of conventional UF membranes. These void-free membranes are more permeable, resulting in high-flux with equivalent or superior product retention. These void-free membranes provide the advantages of fast, high yield processing and smaller systems.

The Pellicon® 2 Mini Ultrafiltration BIOMAX Module feature a filtration area of 0.1 m2 with a pore size of 30 kDa NMWCO.
This product is being discontinued as notified in July 2022. Equivalent and new generation Pellicon™ 3 cassettes are available for ordering. Please contact your local representative or technical support for more information.

Applicazioni

Concentration, desalting or buffer exchange of:
  • Protein solutions
  • Polysaccharide solutions
  • Virus suspensions


Harvest, washing or clarification of:
  • Cell cultures and lysates
  • Colloidal suspensions
  • Viral cultures

Caratteristiche e vantaggi

  • Type A screen (tight screen)
  • Leading-edge void- free membranes to match virtually any separation challenge
  • Short flow path for higher flux and higher resolution separation capability
  • Choice of flow channel configuration providing process optimization capability
  • Predictable, fast, scale-up
  • True linear scalability from laboratory size modules to industrial assemblies for processing thousands of liters
  • Brings higher resolution, improved yields and superior back-pressure resistance
  • Ultracel membranes are composite membranes cast on a microporous substrate for defect-free membranes with superior

Altre note

These products are manufactured in a facility which adheres to FDA Good Manufacturing Practices.
Prefiltration Rquirements: 100 µm

Risultati analitici

Gravimetric Extractables: after a water flush

Note legali

BIOMAX is a registered trademark of Merck KGaA, Darmstadt, Germany
Pellicon is a registered trademark of Merck KGaA, Darmstadt, Germany

Sostituito da

N° Catalogo
Descrizione
Determinazione del prezzo

Codice della classe di stoccaggio

10 - Combustible liquids

Classe di pericolosità dell'acqua (WGK)

WGK 2


Certificati d'analisi (COA)

Cerca il Certificati d'analisi (COA) digitando il numero di lotto/batch corrispondente. I numeri di lotto o di batch sono stampati sull'etichetta dei prodotti dopo la parola ‘Lotto’ o ‘Batch’.

Possiedi già questo prodotto?

I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.

Visita l’Archivio dei documenti

Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..

Contatta l'Assistenza Tecnica.