KVEPA10HH1
Millipore Express® SHR, Opticap® XL Capsule
Opticap® XL 10, cartridge nominal length 10 in. (25 cm)
Sinonimo/i:
Autoclavable Opticap XL 10 Millipore Express SHR single layer 0.1 m 9/16 in. HB/HB
About This Item
Prodotti consigliati
Materiali
polyethersulfone
polypropylene
polypropylene housing
polypropylene support
polypropylene vent cap
silicone seal
Livello qualitativo
Conformità normativa
meets FDA Indirect Food Additive requirements cited in 21 CFR 177-182 (all component materials)
Sterilità
non-sterile
Compatibilità con la sterilizzazione
autoclavable compatible
Nome Commerciale
Opticap® XL 10
Caratteristiche
hydrophilic
Produttore/marchio commerciale
Millipore Express®
Parametri
1.0 bar max. differential pressure (15 psid) at 80 °C (Forward)
1.0 bar max. inlet pressure (15 psi) at 80 °C
100 psi max. differential pressure (6.9 bar) (intermittent)
2.1 bar max. differential pressure (30 psid) at 25 °C (Reverse; intermittent)
2.75 bar max. inlet pressure (40 psi) at 60 °C
25 °C max. inlet temp.
5.5 bar max. differential pressure (80 psid) at 25 °C (Forward)
5.5 bar max. inlet pressure (80 psi) at 23 °C
6.9 bar max. differential pressure (100 psid) at 25 °C (Forward; intermittent)
6.9 bar max. inlet pressure (100 psi) at 23 °C (intermittent)
80 psi max. differential pressure (5.5 bar) (continuous)
80 psig max. inlet pressure
tecniche
sterile filtration: suitable
Lunghezza
33.5 cm (13.2 in.)
Lunghezza nominale della cartuccia
10 in. (25 cm)
Diametro
10.7 cm (4.2 in.)
Superficie filtrante
0.6 m2
Amp. ingresso/ uscita
33.5 cm (13.2 in.)
Diam. connessione uscita
9/16 in.
Impurezze
≤0.25 EU/mL bacterial endotoxins (LAL test, Aqueous extraction)
<0.25 EU/mL USP bacterial endotoxins (LAL test, sample aqueous extraction)
Matrice
Millipore Express® SHR
Dimensione pori
0.1 μm
Punto di bolla
≥2586 mbar (37.5 psig), nitrogen with 70/30% IPA/water at 23 °C
Raccordi
1/4 in. drain/vent hose barb (with double O-ring Seal)
(14 mm (9/16 in.) Hose Barb Inlet and Outlet)
Descrizione generale
Confezionamento
Altre note
- Organism Retention: Microorganism
- Mode of Action: Filtration (size exclusion)
- Application: BioProcessing
- Intended Use: Reduction or removal of microorganism/bioburden
- Instructions for Use: Please refer installation section of OPTICAP® XL Capsules and OPTICAP® XLT Capsules user guide
- Storage Statement: Store in dry location
- Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations.
Nota sulla preparazione
3 autoclave cycles of 60 min @ 126 °C; not in-line steam sterilizable
Risultati analitici
Samples were quantitatively retentive of a minimum Brevundimonas diminuta challenge concentration of 1 x 10⁷ CFU/cm² using ASTM® F838 methodology.
After sterilization and a controlled water flush of 10 L, samples exhibited less than 500 ppb TOC per USP <64> and less than 1.3 µS/cm per USP <64> at 25 °C.
Note legali
Esclusione di responsabilità
Not finding the right product?
Try our Motore di ricerca dei prodotti.
Certificati d'analisi (COA)
Cerca il Certificati d'analisi (COA) digitando il numero di lotto/batch corrispondente. I numeri di lotto o di batch sono stampati sull'etichetta dei prodotti dopo la parola ‘Lotto’ o ‘Batch’.
Possiedi già questo prodotto?
I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.
Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..
Contatta l'Assistenza Tecnica.