Passa al contenuto
Merck
Tutte le immagini(1)

Key Documents

CZRV72TP1

Millipore

Viresolve® NFR Cartridge Filter

Retrovirus removal

Sinonimo/i:

Viresolve NFR Cartridge Filter 20 in. Code 7

Autenticatiper visualizzare i prezzi riservati alla tua organizzazione & contrattuali


About This Item

Codice UNSPSC:
23151806
eCl@ss:
32031690

Materiali

polyethersulfone
polypropylene
silicone seal

Livello qualitativo

Caratteristiche

holdup volume 325 mL

Produttore/marchio commerciale

Viresolve®

Parametri

≤32 mL/min air diffusion at 3.45 bar (50 psig) and 23 °C (in water)
1200-2400 L process volume
3.4 bar max. differential pressure (50 psid) at 25 °C (Reverse)
5.5 bar max. differential pressure (80 psid) at 25 °C (Forward)

tecniche

protein purification: suitable

Lunghezza nominale della cartuccia

20 in. (50 cm)

Superficie filtrante

0.854 m2

Impurezze

≤0.5 EU/mL bacterial endotoxins (LAL test, Aqueous extraction)
<0.5 EU/mL USP bacterial endotoxins (LAL test, sample aqueous extraction)

Estraibili (gravimetria)

≤70 mg/cartridge

Matrice

Viresolve® NFR

Codice cartuccia

Code 7

Cerchi prodotti simili? Visita Guida al confronto tra prodotti

Descrizione generale

Device Configuration: Cartridge

Applicazioni

Retrovirus removal

Confezionamento

Double Easy-Open bag

Altre note

These products are manufactured in a facility which adheres to FDA Good Manufacturing Practices.
Directions for Use

  • Organism Retention: Retrovirus
  • Mode of Action: Filtration (size exclusion)
  • Application: Protein purification
  • Intended Use: Viral clearance
  • Instructions for Use: Please see the user guide (pg. 3) shipped with this product
  • Storage Statement: Store at room temperature
  • Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations.

Nota sulla preparazione

Sterilization Method
After wetting, may be autoclaved for 3 cycles of up to 60 min at 125 °C,

using liquid cycle, slow exhaust.
This product was manufactured with a Viresolve® membrane which meets the criteria for a "non-fiber releasing" filter as defined in 21 CFR 210.3 (b) (6).

Risultati analitici

Gravimetric Extractables: after a 10 minute 1.5 LPM/ft² flush and 24 hours in water at controlled room temperature
Will meet the USP Oxidizable Substances Test requirements after a water flush of ≥8 L

Note legali

VIRESOLVE is a registered trademark of Merck KGaA, Darmstadt, Germany

Certificati d'analisi (COA)

Cerca il Certificati d'analisi (COA) digitando il numero di lotto/batch corrispondente. I numeri di lotto o di batch sono stampati sull'etichetta dei prodotti dopo la parola ‘Lotto’ o ‘Batch’.

Possiedi già questo prodotto?

I documenti relativi ai prodotti acquistati recentemente sono disponibili nell’Archivio dei documenti.

Visita l’Archivio dei documenti

Il team dei nostri ricercatori vanta grande esperienza in tutte le aree della ricerca quali Life Science, scienza dei materiali, sintesi chimica, cromatografia, discipline analitiche, ecc..

Contatta l'Assistenza Tecnica.