HPLC Analysis of Spice Cannabinoids on Ascentis® Express C18

HPLC Analysis of Spice Cannabinoids on Ascentis® Express C18 application for HPLC

Materiais

coluna analítica

Nº do produto
Descrição
Preços

Coluna de HPLC Ascentis® Express C18, 2,7 μm

2.7 μm particle size, L × I.D. 10 cm × 2.1 mm

padrão

Nº do produto
Descrição
Preços

HU-210 solution

100 μg/mL in methanol, (Spice Cannabinoid), ampule of 1 mL, certified reference material, Cerilliant®

JWH-018 solution

100 μg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

JWH-073 solution

100 μg/mL in acetonitrile, ampule of 1 mL, certified reference material, Cerilliant®

JWH-073 N-(3-hydroxybutyl) metabolite solution

100 μg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

CONDITIONS

column

Ascentis Express C18, 10 cm x 2.1 mm I.D., 2.7 μm particles (53823-U)

mobile phase

[A] 5 mM ammonium formate water; [B] 5 mM ammonium formate (95:5, acetonitrile:water)

gradient

68 to 100% B in 8 min; held at 100% B for 2 min

flow rate

0.6 mL/min

pressure

4351 psi (300 bar)

column temp.

30 °C

detector

UV, 265 nm

injection

3 μL

Descrição

Nota de análise

Synthetic cannabinoids (e.g. "Spice") are a type of designer drug used to get a cannabis-type high. New synthetic cannabinoids are continually being introduced as suppliers tweak the molecular structures. The ability to rapidly and reliably identify the continually changing population of these compounds in the blood or urine suspected users is a significant analytical challenge facing forensic chemists. Here is shown the selectivity provided by C18 phase chemistry on an Ascentis Express C18 column with UV detection. Cerilliant CRMs provided reliable identification and quantification.