sample preparation: Solid Phase Extraction
sample/matrix: urine spiked with drugs and metabolites at 300 ng/mL
SPE well plate: Supel-Select HLB 96-well plate, 60 mg/well (575662-U)
condition: 1 mL methanol followed by 1 mL water
sample addition: 1 mL hydrolyzed urine
washing: 1 mL water followed by 1 mL 25% methanol
elution: 1 mL acetonitrile
column: Ascentis Express F5, 10 cm x 2.1 mm I.D., 2.7 μm particles (53569-U)
mobile phase: [A] 5 mM ammonium formate, pH 4.2 (adjusted with formic acid); [B] 5 mM ammonium formate in (95:5 acetonitrile:water) pH 5.2 (adjusted with formic acid)
gradient: 15% B held for 1 min, to 70% B in 2 min, held at 70% B for 7 min
flow rate: 0.4 mL/min
column temp.: 35 °C
detector: ESI(+) TOF, full scan, m/z 100-1000
injection: 3 μL
sample: 300 ng/mL
adequação
application for HPLC
application for SPE
aplicação(ões)
clinical
forensics and toxicology
Nota de análise
Informações legais
Analito
Padrão
coluna analítica
tubo ou placa SPE
Escolha uma das versões mais recentes:
Certificados de análise (COA)
Lamentamos, não temos COA para este produto disponíveis online no momento.
Se precisar de ajuda, entre em contato Atendimento ao cliente
Já possui este produto?
Encontre a documentação dos produtos que você adquiriu recentemente na biblioteca de documentos.
Nossa equipe de cientistas tem experiência em todas as áreas de pesquisa, incluindo Life Sciences, ciência de materiais, síntese química, cromatografia, química analítica e muitas outras.
Entre em contato com a assistência técnica