G005814

GC Analysis of FAMEs in Atlantic Salmon on Omegawax

suitable for GC

Faça loginpara ver os preços organizacionais e de contrato


About This Item

técnica(s)

gas chromatography (GC): suitable

parâmetros de teste

column: Omegawax, 30 m x 0.25 mm I.D., 0.25μm (24136)
oven: 170 °C, 1 °C/min to 225°C
inj. temp.: 250 °C
detector: FID, 260 °C
carrier gas: helium, 1.2 mL/min
injection: 1 μL, 100:1 split
liner: 4 mm ID FocusLiner with taper
sample: 1. Fish Oil from capsule, prepared according to AOAC Method 991.39 2. Atlantic, farm-raised salmon, prepared according to AOCS Official Method Ce 1k-09

aplicação(ões)

food and beverages

Nota de análise

Samples of farm-raised Atlantic salmon were prepared and analyzed for fatty acids. The salmon was extracted using an acid digestion /alkali hydrolysis followed by methylation as described in AOCS Official Method Ce 1k-09 (2). The resulting extracts were then analyzed by GC/FID on an Omegawax 250 capillary column as described in AOCS Official Method Ce 1i-07.

Informações legais

coluna analítica

Nº do produto
Descrição
Preços

Escolha uma das versões mais recentes:

Certificados de análise (COA)

Lot/Batch Number

Lamentamos, não temos COA para este produto disponíveis online no momento.

Se precisar de ajuda, entre em contato Atendimento ao cliente

Já possui este produto?

Encontre a documentação dos produtos que você adquiriu recentemente na biblioteca de documentos.

Visite a Biblioteca de Documentos

Nossa equipe de cientistas tem experiência em todas as áreas de pesquisa, incluindo Life Sciences, ciência de materiais, síntese química, cromatografia, química analítica e muitas outras.

Entre em contato com a assistência técnica