G005711
LC/MS Analysis of Omeprazole and Metabolites in Plasma on Ascentis® Express RP Amide after Sample Prep using HybridSPE®-Phospholipid
application for HPLC, application for SPE
About This Item
técnica(s)
HPLC: suitable
solid phase extraction (SPE): suitable
parâmetros de teste
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: rat plasma spiked with omeprazole, phenacetin and metabolites at 2 μg/mL
SPE well plate: HybridSPE-Phospholipid, 96-well plate, 15 mg/0.8 mL (575656-U)
sample addition: 100 μL plasma to each well, followed by 300 mL of 1% formic acid in acetonitrile
elution: apply vacuum
eluate post-treatment: eluate was collected, evaporated, and reconstituted in mobile phase.
column: Ascentis Express RP-Amide, 5 cm x 2.1 mm I.D., 2.7 μm particles (53911-U)
mobile phase: [A] water, 10 mM ammonium formate, pH 3.4; [B] methanol, 10 mM ammonium formate, pH 3.4; (65:35, A:B)
flow rate: 0.3 mL/min
column temp.: 35 °C
detector: ESI(+), m/z = 50-400
injection: 1 μL
adequação
application for HPLC
application for SPE
aplicação(ões)
clinical
Aplicação
Nota de análise
Informações legais
Analito
Padrão
coluna analítica
tubo ou placa SPE
Escolha uma das versões mais recentes:
Certificados de análise (COA)
It looks like we've run into a problem, but you can still download Certificates of Analysis from our Documentos section.
Se precisar de ajuda, entre em contato Atendimento ao cliente
Já possui este produto?
Encontre a documentação dos produtos que você adquiriu recentemente na biblioteca de documentos.
Nossa equipe de cientistas tem experiência em todas as áreas de pesquisa, incluindo Life Sciences, ciência de materiais, síntese química, cromatografia, química analítica e muitas outras.
Entre em contato com a assistência técnica