Pular para o conteúdo
Merck
Todas as fotos(1)

Key Documents

G005674

HPLC Analysis of Benzodiazepines on Ascentis® Express Phenyl-Hexyl, Methanol Mobile Phase

application for HPLC

Faça loginpara ver os preços organizacionais e de contrato


About This Item

técnica(s)

HPLC: suitable

parâmetros de teste

column: Ascentis Express Phenyl-Hexyl, 5 cm x 4.6 mm I.D., 2.7 μm particles (53348-U)
mobile phase: [A] water; [B] methanol; (35:65, A:B)
flow rate: 1.4 mL/min
pressure: 3600 psi (248 bar)
column temp.: 35 °C
detector: UV, 250 nm
injection: 2 μL
sample: 0.1 g/L in 20% methanol

adequação

application for HPLC

aplicação(ões)

clinical
pharmaceutical (small molecule)

Nota de análise

The Phenyl-Hexyl phase of the Ascentis Express product line displays the best selectivity of this sample; selectivity is markedly affected by choice of organic modifier. The application also demonstrates feasibility of fast LC with Ascentis Express by virtue of low backpressure and high efficiency.

Informações legais

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

Analito

Descrição

    Clonazepam
    Oxazepam
    Alprazolam
    Desmethyldiazepam
    Diazepam

Escolha uma das versões mais recentes:

Certificados de análise (COA)

Lot/Batch Number

Lamentamos, não temos COA para este produto disponíveis online no momento.

Se precisar de ajuda, entre em contato Atendimento ao cliente

Já possui este produto?

Encontre a documentação dos produtos que você adquiriu recentemente na biblioteca de documentos.

Visite a Biblioteca de Documentos

Nossa equipe de cientistas tem experiência em todas as áreas de pesquisa, incluindo Life Sciences, ciência de materiais, síntese química, cromatografia, química analítica e muitas outras.

Entre em contato com a assistência técnica