G004366
HPLC Analysis of Verapamil and Metabolites in Rat Plasma on Ascentis® Express C18 after Phospholipid Removal using HybridSPE®-Phospholipid
application for SPE, application for LC-MS
About This Item
técnica(s)
LC/MS: suitable
solid phase extraction (SPE): suitable
parâmetros de teste
SPE well plate: HybridSPE-Precipitation 96-well Plate, 50 mg/well (575656-U)
sample preparation: SPE (Solid Phase Extraction)
column: Ascentis Express C18, 5 cm x 2.1 mm I.D., 2.7 μm particles (53822-U)
sample/matrix: Rat plasma spiked with verapamil and metabolites at 10 ng/mL
mobile phase: [A] water; [B] acetonitrile, pH 2.55 adjust with formic acid (30:70, A:B)
flow rate: 0.6 mL/min
column temp.: 35 °C
detector: TOF/MS
injection: 1 μL
sample addition: 100 μL spiked rat plasma followed by 300 μL 1% formic acid in acetonitrile
elution: apply vacuum
adequação
application for LC-MS
application for SPE
aplicação(ões)
clinical
Informações legais
tubo ou placa SPE
Choose from one of the most recent versions:
Certificados de análise (COA)
Sorry, we don't have COAs for this product available online at this time.
If you need assistance, please contact Atendimento ao cliente
Já possui este produto?
Encontre a documentação dos produtos que você adquiriu recentemente na biblioteca de documentos.
Nossa equipe de cientistas tem experiência em todas as áreas de pesquisa, incluindo Life Sciences, ciência de materiais, síntese química, cromatografia, química analítica e muitas outras.
Entre em contato com a assistência técnica