G003758
GC Analysis of Pesticides in Spinach on Equity-1 after SPE using Supelclean™ ENVI-Carb-II/PSA
suitable for GC, application for SPE
About This Item
técnica(s)
gas chromatography (GC): suitable
solid phase extraction (SPE): suitable
parâmetros de teste
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: 10 g of fresh spinach spiked at 0.2 ppm each pesticide
SPE tube/cartridge: Supelclean ENVI-Carb-II/PSA, 500mg/300mg/6mL (55119-U)
condition: 5 mL acetonitrile:toluene (3:1, v/v)
sample addition: 1 mL concentrated sample extract
elution: 20 mL acetonitrile:toluene (3:1, v/v)
eluate post-treatment: Evaporate eluate to dryness with nitrogen at 40 °C. Reconstitute in 1 mL acetone:hexane (1:1, v/v).
column: Equity-1, 30 m x 0.25 mm I.D., 0.25 μm (28046-U)
oven: 50 °C (5 min), 25 °C/min to 125 °C, 10 °C/min to 30 °C (8 min)
inj. temp.: 200 °C, aux: 325 °C (8 min)
detector: MSD, scan range 45-450 amu
carrier gas: helium, 0.9 mL/min., constant flow mode
injection: 1 μL, splitless (splitter open at 1 min)
liner: 4 mm I.D., single taper
adequação
application for SPE
aplicação(ões)
agriculture
food and beverages
Aplicação
Homogenize 10 g of spinach in 10 mL of acetonitrile.
Informações legais
coluna analítica
tubo ou placa SPE
Escolha uma das versões mais recentes:
Certificados de análise (COA)
Lamentamos, não temos COA para este produto disponíveis online no momento.
Se precisar de ajuda, entre em contato Atendimento ao cliente
Já possui este produto?
Encontre a documentação dos produtos que você adquiriu recentemente na biblioteca de documentos.
Nossa equipe de cientistas tem experiência em todas as áreas de pesquisa, incluindo Life Sciences, ciência de materiais, síntese química, cromatografia, química analítica e muitas outras.
Entre em contato com a assistência técnica