Pular para o conteúdo
Merck
Todas as fotos(1)

Documentos Principais

G001645

GC Analysis of Trace Odors in Drinking Water on Equity-5 after SPME using 50/30 μm DVB/Carboxen/PDMS Fiber

suitable for GC, application for SPME

Faça loginpara ver os preços organizacionais e de contrato


About This Item

técnica(s)

gas chromatography (GC): suitable
solid phase microextraction (SPME): suitable

parâmetros de teste

sample/matrix: 25 mL of water containing 25% NaCl and drinking water odors kit (47529-U)
SPME fiber: 2 cm StableFlex coated with 50/30 μm DVB/Carboxen/PDMS (57348-U)
extraction: headspace, 65 °C (30 min)
desorption process: 3 min at 260 °C
column: Equity-5, 30 m × 0.25 mm, 0.25 μm (28089-U)
oven: 60 °C (2 min) to 200 °C at 8 °C/min
flow rate: helium, 37 cm/sec @ 60 °C (1 mL/min constant flow)
inj. temp.: SPME fiber, splitless opened at after 1 min at 50 mL/min
detector: MS, selected ions (SIM) 95, 112, 124, 137, 197; interface at 280 °C
liner: 0.75 mm SPME liner

adequação

application for SPME

aplicação(ões)

environmental
food and beverages

Informações legais

Analito

Descrição

    2-Methylisoborneol
    (±)-Geosmin
    2-Isopropyl-3-methoxypyrazine
    2,4,6-Trichloroanisole
    2-Isobutyl-3-methoxypyrazine

Escolha uma das versões mais recentes:

Certificados de análise (COA)

Lot/Batch Number

Lamentamos, não temos COA para este produto disponíveis online no momento.

Se precisar de ajuda, entre em contato Atendimento ao cliente

Já possui este produto?

Encontre a documentação dos produtos que você adquiriu recentemente na biblioteca de documentos.

Visite a Biblioteca de Documentos

Nossa equipe de cientistas tem experiência em todas as áreas de pesquisa, incluindo Life Sciences, ciência de materiais, síntese química, cromatografia, química analítica e muitas outras.

Entre em contato com a assistência técnica