G001056
HPLC Analysis of Barbiturates in Serum on Discovery® C18 after SPE using Discovery® DSC-18Lt
application for SPE, application for HPLC
About This Item
técnica(s)
HPLC: suitable
solid phase extraction (SPE): suitable
parâmetros de teste
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: 0.5 mL porcine serum spiked with 0.5 μg/mL or 1.0 μg/mL each analyte then diluted with 0.5 mL water
SPE tube/cartridge: Discovery DSC-18Lt, 500 mg/3 mL (52613-U)
condition: 2 mL methanol; 2 mL DI water
sample addition: 1 mL at 0.75 mL/min
washing: 2 mL 5% methanol, then vacuum or air dry for 5-10 min
elution: 1-2 mL methanol
eluate post-treatment: dry eluate with nitrogen purge (40 °C; 15-20 min), reconstitute in 20 μL mobile phase
column: Discovery C18, 15 cm × 4.6 mm, 5 μm preceded by a 2 cm C18 guard column and 0.5 μm frit filter (504955)
mobile phase: [A] methanol; [B] water (40:60, A:B)
flow rate: 1 mL/min
column temp.: 30 °C
detector: UV, 214 nm
injection: 30 μL, diluted porcine serum extract
adequação
application for HPLC
application for SPE
aplicação(ões)
forensics and toxicology
Nota de análise
Informações legais
Analito
Padrão
coluna analítica
tubo ou placa SPE
Escolha uma das versões mais recentes:
Certificados de análise (COA)
It looks like we've run into a problem, but you can still download Certificates of Analysis from our Documentos section.
Se precisar de ajuda, entre em contato Atendimento ao cliente
Já possui este produto?
Encontre a documentação dos produtos que você adquiriu recentemente na biblioteca de documentos.
Nossa equipe de cientistas tem experiência em todas as áreas de pesquisa, incluindo Life Sciences, ciência de materiais, síntese química, cromatografia, química analítica e muitas outras.
Entre em contato com a assistência técnica