About This Item
Produtos recomendados
técnica(s)
protein clean-up: suitable
aplicação(ões)
sample preparation
Descrição geral
The Chemicon Detergent-OUT Kit is a simple high performance method for removing detergents from protein solutions. Removes detergents without significant dilution of protein solution. Simply load the protein solution on the Detergent-OUT Spin column and collect detergent free protein solution. The Med column is suitable for removing detergents from a 1.0 mL solution of 1.5% Triton-X100 detergents (or an equivalent of 15 μL Triton-X100 or 15 mg detergents), provided that both the columns and proteins solutions are prepared in water.
Aplicação
PROTOCOL:
1. Invert the column several times to re-suspend the resin material. Remove the tip of the column and let the liquid drain off.
2. Buffer Equilibration: Detergent-OUT columns are supplied in deionized water. A better result is obtained when the column and protein solution are in water or phosphate buffer. Columns may be equilibrated with other types of buffer. Note: Do not equilibrate the columns with Tris buffer and other such buffers. Apply about 1.0 mL of a desired buffer into the column. Let the buffer drain into collection tube. Repeat this process 3 times.
3. Place the column in a 15mL centrifuge tube. Centrifuge at 1000 xg for 20-30 seconds. Discard the liquid collected in the centrifuge tube.
4. Place the column back in the same centrifuge tube. Carefully apply the protein suspension to the center of the column. Wait for 5 minutes..
5. After loading the column centrifuge at 1000 xg for 20-30 seconds. Collect the liquid comprising purified protein material free from detergent originally present in the sample.
1. Invert the column several times to re-suspend the resin material. Remove the tip of the column and let the liquid drain off.
2. Buffer Equilibration: Detergent-OUT columns are supplied in deionized water. A better result is obtained when the column and protein solution are in water or phosphate buffer. Columns may be equilibrated with other types of buffer. Note: Do not equilibrate the columns with Tris buffer and other such buffers. Apply about 1.0 mL of a desired buffer into the column. Let the buffer drain into collection tube. Repeat this process 3 times.
3. Place the column in a 15mL centrifuge tube. Centrifuge at 1000 xg for 20-30 seconds. Discard the liquid collected in the centrifuge tube.
4. Place the column back in the same centrifuge tube. Carefully apply the protein suspension to the center of the column. Wait for 5 minutes..
5. After loading the column centrifuge at 1000 xg for 20-30 seconds. Collect the liquid comprising purified protein material free from detergent originally present in the sample.
Embalagem
10 columns
Componentes
Detergent-OUT Columns-DTG-100X 10 Med
Armazenamento e estabilidade
Store at 2-8°C for up to 6 months from date of purchase. To avoid possible microbial contamination, dispense all solutions aseptically.
Exoneração de responsabilidade
Unless otherwise stated in our catalog or other company documentation accompanying the product(s), our products are intended for research use only and are not to be used for any other purpose, which includes but is not limited to, unauthorized commercial uses, in vitro diagnostic uses, ex vivo or in vivo therapeutic uses or any type of consumption or application to humans or animals.
Certificados de análise (COA)
Busque Certificados de análise (COA) digitando o Número do Lote do produto. Os números de lote e remessa podem ser encontrados no rótulo de um produto após a palavra “Lot” ou “Batch”.
Já possui este produto?
Encontre a documentação dos produtos que você adquiriu recentemente na biblioteca de documentos.
Nossa equipe de cientistas tem experiência em todas as áreas de pesquisa, incluindo Life Sciences, ciência de materiais, síntese química, cromatografia, química analítica e muitas outras.
Entre em contato com a assistência técnica