GC Analysis of Polynuclear Aromatic Hydrocarbons (PAHs) in Grilled Hamburger on SPB®-608 (20 m x 0.18 mm I.D., 0.18 μm) after QuEChERS Cleanup using Supel QuE Z-Sep, Fast GC Analysis

suitable for GC, application for SPE

Se connecterpour consulter vos tarifs contractuels et ceux de votre entreprise/organisme


About This Item

Technique(s)

gas chromatography (GC): suitable
solid phase extraction (SPE): suitable

Paramètres du test

sample preparation: Dispersive (QuEChERS)
sample/matrix: grilled hamburger spiked with PAHs at 100 ng/g
extraction tube: Sodium chloride/magnesium sulfate (55294-U)
extraction process: add 25 mL acetonitrile, contents of unbuffered salt tube (55294-U); shake for 5 minutes; centrifuge for 5 minutes at 3200 rpm
clean-up tube: Supel QuE Z-Sep tube (55403-U)
clean-up process: transfer 3 mL of acetonitrile layer into a Supel QuE Z-Sep cleanup tube (55403-U); shake 1 minute; centrifuge for 3 minutes at 3400 rpm; remove 1 mL supernatant for GC-MS analysis
column: SPB-608, 20 m x 0.18 mm I.D. x 0.18 μm
oven: 60 °C (1 min), 25 °C/min to 275 °C, 10 °C/min to 300 °C (13 min)
inj. temp.: 265 °C
detector: MS-SIM
MSD interface: 300 °C
carrier gas: helium, 1.5 mL/min constant flow
injection: 1 μL splitless (1 min.)
liner: 4 mm I.D. FocusLiner with taper
sample: grilled hamburger extract

Adéquation

application for SPE

Application(s)

food and beverages

Remarque sur l'analyse

Sample pretreatment: Place 5 g ground hamburger sample into 50 mL centrifuge tube (55248-U). Spike with 100 ng/g target PAHs and internal standards. Let sit 10 min. Add 5 mL water and homogenize.
This application demonstrates the use of dispersive SPE or QuEChERS for the extraction and cleanup of grilled hamburger for GC-MS analysis of polynuclear aromatic hydrocarbons (PAHs). Z-Sep sorbent was used in the cleanup step and found to reduce background and yield recoveries >80% for a majority of the PAHs tested.

Autres remarques

Informations légales

FocusLiner is a trademark of SGE Analytical Science Pty Ltd.
SPB is a registered trademark of Merck KGaA, Darmstadt, Germany
Supel is a trademark of Sigma-Aldrich Co. LLC

Tube ou plaque pour SPE

Réf. du produit
Description
Tarif

Faites votre choix parmi les versions les plus récentes :

Certificats d'analyse (COA)

Lot/Batch Number

Désolés, nous n'avons pas de COA pour ce produit disponible en ligne pour le moment.

Si vous avez besoin d'assistance, veuillez contacter Service Clients

Déjà en possession de ce produit ?

Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.

Consulter la Bibliothèque de documents

Notre équipe de scientifiques dispose d'une expérience dans tous les secteurs de la recherche, notamment en sciences de la vie, science des matériaux, synthèse chimique, chromatographie, analyse et dans de nombreux autres domaines..

Contacter notre Service technique