HPLC Analysis of Furans on Ascentis® Express C18

application for HPLC

Se connecterpour consulter vos tarifs contractuels et ceux de votre entreprise/organisme


About This Item

Technique(s)

HPLC: suitable

Paramètres du test

column: Ascentis Express C18, 5 cm x 3.0 mm I.D., 2.7 μm particles (53811-U)
mobile phase: [A] water; [B] acetonitrile; (80:20, A:B)
flow rate: 0.4 mL/min
column temp.: 35 °C
detector: UV, 230 nm
injection: 1 μL
sample: 20 μg/mL in 80:20, water: acetonitrile

Adéquation

application for HPLC

Application(s)

chemicals and industrial polymers

Remarque sur l'analyse

This application demonstrates the suitability of Ascentis Express C18 for the efficient separation of furans as listed in ASTM method 05837-99.

Informations légales

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

Analyte

Description

    5-Hydroxymethyl-2-furaldehyde
    Furfuryl alcohol
    Furfural
    2-Furyl methyl ketone
    5-Methyl-2-furaldehyde

Colonne analytique

Étalon

Réf. du produit
Description
Tarif

Faites votre choix parmi les versions les plus récentes :

Certificats d'analyse (COA)

Lot/Batch Number

Désolés, nous n'avons pas de COA pour ce produit disponible en ligne pour le moment.

Si vous avez besoin d'assistance, veuillez contacter Service Clients

Déjà en possession de ce produit ?

Retrouvez la documentation relative aux produits que vous avez récemment achetés dans la Bibliothèque de documents.

Consulter la Bibliothèque de documents

Notre équipe de scientifiques dispose d'une expérience dans tous les secteurs de la recherche, notamment en sciences de la vie, science des matériaux, synthèse chimique, chromatographie, analyse et dans de nombreux autres domaines..

Contacter notre Service technique