HPLC Analysis of Fat Soluble Vitamins by Normal Phase Chromatography on Ascentis® Si (25 cm Column)

application for HPLC

Anmeldenzur Ansicht organisationsspezifischer und vertraglich vereinbarter Preise


About This Item

Methode(n)

HPLC: suitable

Testparameter

column: Ascentis Si, 15 cm x 4.6 mm I.D., 5 μm particles (581512-U)
mobile phase: [A] hexane; [B] ethyl acetate
gradient: 10 to 30% B in 10 min; held at 30% B for 2 min
flow rate: 1.0 mL/min
column temp.: 30 °C
detector: UV, 290 nm
injection: 10 μL

Eignung

application for HPLC

Anwendung(en)

vitamins, nutraceuticals, and natural products

Hinweis zur Analyse

Ascentis silica is used in the analysis of fat-soluble vitamins using a normal phase gradient. These include vitamin K 3 (menadione), vitamin E (alpha and gamma tocopherol), and vitamin D3 (cholecalciferol).

Rechtliche Hinweise

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

Analyt

Beschreibung

    (±)-α-Tocopherol, analytical standard
    Menadione (K3), analytical standard
    (+)-γ-Tocopherol, analytical standard
    Cholecalciferol (D3), analytical standard

Analytische Säule

Produkt-Nr.
Beschreibung
Preisangaben

Standard

Produkt-Nr.
Beschreibung
Preisangaben

Hier finden Sie alle aktuellen Versionen:

Analysenzertifikate (COA)

Lot/Batch Number

Leider sind derzeit keine COAs für dieses Produkt online verfügbar.

Wenn Sie Hilfe benötigen, wenden Sie sich bitte an Kundensupport

Besitzen Sie dieses Produkt bereits?

In der Dokumentenbibliothek finden Sie die Dokumentation zu den Produkten, die Sie kürzlich erworben haben.

Die Dokumentenbibliothek aufrufen

Unser Team von Wissenschaftlern verfügt über Erfahrung in allen Forschungsbereichen einschließlich Life Science, Materialwissenschaften, chemischer Synthese, Chromatographie, Analytik und vielen mehr..

Setzen Sie sich mit dem technischen Dienst in Verbindung.