Direkt zum Inhalt
Merck

KVSCA10TT1

Millipore

Multimedia Durapore® 1.2/0.22 µm, Opticap® XL Capsule

Opticap® XL 10, cartridge nominal length 10 in. (25 cm), filter diam. 4.2 in. (10.7 cm)

Anmeldenzur Ansicht organisationsspezifischer und vertraglich vereinbarter Preise


About This Item

UNSPSC-Code:
23151806

Materialien

mixed cellulose esters (MCE) prefilter
polypropylene filter
polypropylene housing
polypropylene support
silicone seal

Qualitätsniveau

Einhaltung gesetzlicher Vorschriften

FDA 21CFR177-182 (all component materials)

Sterilität

non-sterile

Kompatibilität mit Sterilisationsverfahren

autoclavable compatible

Produktlinie

Opticap® XL 10

Leistungsmerkmale

hydrophilic

Hersteller/Markenname

Opticap®

Parameter

≤10.8 mL/min air diffusion at 2.75 bar (40 psig) and 23 °C (in water)
1 bar max. inlet pressure (15 psi) at 80 °C
1.0 bar max. differential pressure (15 psid) at 80 °C (Forward:)
2.8 bar max. inlet pressure (40 psi) at 60 °C
25 °C max. inlet temp.
3.5 bar max. differential pressure (50 psid) at 25 °C (Reverse; intermittent)
5.5 bar max. differential pressure (80 psid) at 25 °C (Forward:)
5.5 bar max. inlet pressure (80 psi) at 25 °C
80 psig max. inlet pressure

Methode(n)

sterile filtration: suitable

Breite

4.2 in.

Nennlänge Kartusche

10 in. (25 cm)

device L

33.5 cm (13.2 in.)

device size

10 in.

Filterdurchm.

4.2 in. (10.7 cm)

Filtrationsfläche

0.56 m2

Durchm. Einlassanschluss

1-1/2 in.

Durchm. Auslassanschluss

1-1/2 in.

Verunreinigungen

≤0.5 EU/mL bacterial endotoxins (LAL test, aqueous extraction)
<0.5 EU/mL bacterial endotoxins (LAL test, sample aqueous extraction)

Matrix

Multimedia Durapore®

pore size

1.2/0.22 μm pore size

Blasenpunkt

≥3450 mbar (50 psig), air with water at 23 °C

Fitting

38 mm (1 1/2 in.) fitting (sanitary flange inlet/oulet)
1/4 in. fitting (vent/drain hose barb with double O-ring seal)
(38 mm (1 1/2 in.) Sanitary Flange Inlet and Outlet)

Suchen Sie nach ähnlichen Produkten? Aufrufen Leitfaden zum Produktvergleich

Allgemeine Beschreibung

Device Configuration: Capsule

Verpackung

Double Easy-Open bag

Angaben zur Herstellung

Sterilization Method:
3 autoclave cycles of 60min at 3°C; not in-line steam sterilizable

This product was manufactured with a Durapore® membrane which meets the criteria for a "non-fiber releasing" filter as defined in 21CFR210.3(b)(6).

Hinweis zur Analyse

Bacterial Retention: Samples were quantitatively retentive of a minimum Brevundimonas diminuta challenge concentration of 1 x 107 CFU/cm2 using ASTM® F838 methodology.
Gravimetric Extractables: ≤ 75 mg per capsule after 24 hours in water at controlled room temperature.
Will meet the USP Oxidizable Substances Test requirements after a water flush of ≥ 5 L

Sonstige Hinweise

Directions for Use:

  • Organism Retention: Microorganism
  • Mode of Action: Filtration (size exclusion)
  • Application: BioProcessing
  • Intended Use: Reduction or removal of microorganism/bioburden
  • Instructions for Use: Please refer Wetting Instructions for Filter Units with Durapore® Membrane user guide
  • Storage Statement: Store in dry location
  • Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations

Rechtliche Hinweise

ASTM is a registered trademark of American Society for Testing and Materials
Durapore is a registered trademark of Merck KGaA, Darmstadt, Germany
OPTICAP is a registered trademark of Merck KGaA, Darmstadt, Germany

Sie haben nicht das passende Produkt gefunden?  

Probieren Sie unser Produkt-Auswahlhilfe. aus.

Lagerklassenschlüssel

4.1B - Flammable solid hazardous materials

WGK

WGK 3


Analysenzertifikate (COA)

Suchen Sie nach Analysenzertifikate (COA), indem Sie die Lot-/Chargennummer des Produkts eingeben. Lot- und Chargennummern sind auf dem Produktetikett hinter den Wörtern ‘Lot’ oder ‘Batch’ (Lot oder Charge) zu finden.

Besitzen Sie dieses Produkt bereits?

In der Dokumentenbibliothek finden Sie die Dokumentation zu den Produkten, die Sie kürzlich erworben haben.

Die Dokumentenbibliothek aufrufen

Unser Team von Wissenschaftlern verfügt über Erfahrung in allen Forschungsbereichen einschließlich Life Science, Materialwissenschaften, chemischer Synthese, Chromatographie, Analytik und vielen mehr..

Setzen Sie sich mit dem technischen Dienst in Verbindung.