Saltar al contenido
MilliporeSigma

US EPA Method 632: HPLC Analysis of Pesticides on Ascentis® C18

US EPA Method 632: HPLC Analysis of Pesticides on Ascentis® C18 application for HPLC

Materiales

analytical column

Referencia del producto
Descripción
Precios

Ascentis® C18 HPLC Column

5 μm particle size, L × I.D. 25 cm × 4.6 mm

Ascentis® C8 HPLC Column

3 μm particle size, L × I.D. 15 cm × 4.6 mm

Ascentis® C8 HPLC Column

5 μm particle size, L × I.D. 25 cm × 4.6 mm

Columna para HPLC Ascentis® Express C18, 2,7 μm

2.7 μm particle size, L × I.D. 15 cm × 4.6 mm

Ascentis® Express C8 HPLC Column

2.7 μm particle size, L × I.D. 15 cm × 4.6 mm

For use with

Referencia del producto
Descripción
Precios

Wheaton wide-mouth bottle

capacity 32 oz (1000 mL), white polypropylene cap, PTFE flat liner, amber

Holders para Supelguard precolumnas

Stand-alone, for use with Supelguard cartridges (2 cm L. x 2.1 to 4.0 mm I.D.)

Guard cartridge

Referencia del producto
Descripción
Precios

Ascentis® C18 Supelguard Guard Cartridge

5 μm particle size, L × I.D. 2 cm × 4 mm, pkg of 2 ea

Ascentis® C18 Supelguard Guard Cartridge

3 μm particle size, L × I.D. 2 cm × 4 mm, pkg of 2 ea

Ascentis® C8 Supelguard Guard Cartridge

5 μm particle size, L × I.D. 2 cm × 4 mm, pkg of 2 ea

CONDITIONS

column

Ascentis C18, 25 cm x 4.6 mm I.D., 5 μm particles (581325-U)

mobile phase

[A] water; [B] acetonitrile

gradient

10 to 50% B in 13 min; to 60% B in 10 min; to 90% B in 6 min; held for 5 min

flow rate

1.0 mL/min

column temp.

30 °C

detector

UV, 210 nm

injection

10 μL

sample

as indicated in 16% acetonitrile in water

Descripción

Legal Information

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany