HPLC Analysis of Sugars and Taurine in an Energy Drink on Ascentis® Express HILIC using UV and ELSD Detection in Series

HPLC Analysis of Sugars and Taurine in an Energy Drink on Ascentis® Express HILIC using UV and ELSD Detection in Series application for HPLC

CONDITIONS

column

Ascentis Express HILlC, 10 cm x 3.0 mm I.D., 2.7 μm particles (53970-U)

mobile phase

[A] 100 mM ammonium acetate, pH 5.0 with acetic acid; [B] water; [C] acetonitrile; (09:01:90, A:B:C)

flow rate

0.6 mL/min

pressure

812 psi (56 bar)

column temp.

35 °C

detector

ELSD, 55 °C, 3.5 bar nitrogen

injection

2 μL

sample

dilute1:9 in acetonitrile

Descripción

Analysis Note

This application demonstrates the suitability of Ascentis Express HILIC for the efficient separation of common drink additives used in caffeinated energy drinks. The analysis was performed with UV and ELSD detectors in series, this chromatogram being from the ELSD.