Saltar al contenido
MilliporeSigma

HPLC Analysis of Cortocisteroids in Urine on Discovery® HS F5 after SPE using Discovery® DSC-CN

HPLC Analysis of Cortocisteroids in Urine on Discovery® HS F5 after SPE using Discovery® DSC-CN application for SPE, application for HPLC

Materiales

analytical column

Referencia del producto
Descripción
Precios

Columna para HPLC Discovery® HS F5

3 μm particle size, L × I.D. 5 cm × 4.6 mm

CONDITIONS

sample preparation

SPE (Solid Phase Extraction)

sample/matrix

human urine spiked with corticosteroids at 0.5 and 1.0 μg/mL and diluted in DI water (1:1, v/v)

SPE well plate

Discovery DSC-CN 96-well SPE, 100 mg/well (575636-U)

condition

1 mL methanol; 1 mL DI water.

sample addition

1 mL spiked urine

washing

1 mL 20% methanol

elution

1 mL methanol

eluate post-treatment

Evaporate eluate with nitrogen at 30 °C. Reconstitute in 200 μL mobile phase.

column

Discovery HS F5, 5 cm x 4.6 mm I.D., 3 μm particles (567504-U)

mobile phase

[A] methanol; [B] DI water; (40:60, A:B)

flow rate

1.5 mL/min

column temp.

35 °C

detector

UV, 240 nm

injection

5 μL

Descripción

Legal Information

Discovery is a registered trademark of Merck KGaA, Darmstadt, Germany