LC/MS/MS Analysis of Aflatoxins in Cannabis on Ascentis® Express 2.7 μm Phenyl-Hexyl after Cleanup using Supel Tox AflaZea SPE

LC/MS/MS Analysis of Aflatoxins in Cannabis on Ascentis® Express 2.7 μm Phenyl-Hexyl after Cleanup using Supel™ Tox AflaZea SPE application for LC-MS, application for SPE

材料

分析柱

產品號碼
描述
訂價

Ascentis® Express Phenyl-Hexyl,2.7μm HPLC色谱柱

2.7 μm particle size, L × I.D. 5 cm × 2.1 mm

標準

產品號碼
描述
訂價

黄曲霉毒素混合物 4 溶液

analytical standard

附屬品

產品號碼
描述
訂價

黑色塑料盖

Solid top, white rubber liner, thread for 8-425, pkg of 1000 ea

Visiprep SPE 真空固相萃取装置

DL (Disposable Liner), 12-port model

用于Visiprep DL防交叉污染SPE装置(包含在57044和57265)的一次性小柱连接管

PTFE, pk of 100

经硅烷化处理螺纹口样品瓶

volume 2 mL, amber glass vial, O.D. × H × I.D. 12 mm × 32 mm × 4.6 mm

QuEChERS Shaker and Rack Starter Kit

AC/DC input 115 V AC (USA Compatible Plug)

CONDITIONS

sample preparation

Solid Phase Extraction

sample/matrix

cannabis extract spiked with aflatoxins at corresponding concentrations of 6 ppb each (for Aflatoxin B2 and G2) and 24 ppb each (for Afltoxin B1 and G1) in cannabis

SPE tube/cartridge

Supel Tox AflaZea SPE, 6 mL (55314-U)

sample addition

2 mL of sample extract

elution

apply strong vacuum and collect the sample into plastic tube

eluate post-treatment

dilute 0.2 mL of the extract with 0.8 mL distilled water and mix well

column

Ascentis Express Phenyl-Hexyl, 5 cm x 2.1 mm I.D., 2.7 μm particles (53334-U)

mobile phase

[A] 5 mM ammonium formate with 1% formic acid in water; [B] 5 mM ammonium formate with 1% formic acid in methanol

gradient

30% B to 60% B in 3 min; to 100% B in 2 in; held at 100% B for 2 min; to 30% B in 0.1 min; held at 30% B for 1.4 min

flow rate

0.4 mL/min

pressure

5511 psi (380 bar)

column temp.

40 °C

detector

MS, ESI(+), MRM 331.3/189.0, 329.1/243.0, 315.9/259.0, 313.1/241.0

injection

10 μL

描述

分析報告

The complex composition of cannabis makes it difficult to apply sample cleanup methods commonly used for other commodities. In this application, interference removal Supel Tox AflaZea SPE was used for sample cleanup providing quick and effective sample preparation prior to mass spectrometric analysis. Sample comprised 0.5 g of finely ground cannabis mixed with 10 mL of 84:16 acetonitrile:water for 30 minutes and centrifuged. It is recommended to use silanized glass vials for this analysis.