跳轉至內容
Merck

HPLC Analysis of Warfarin Enantiomers on Astec® CYCLOBOND I 2000 (UV at 254 nm)

HPLC Analysis of Warfarin Enantiomers on Astec® CYCLOBOND I 2000 (UV at 254 nm) application for HPLC

材料

分析柱

產品號碼
描述
訂價

Astec® CYCLOBOND I 2000 手性 HPLC 色谱柱

5 μm particle size, L × I.D. 25 cm × 4.6 mm

CONDITIONS

column

CYCLOBOND I 2000, 25 cm × 4.6 mm I.D., 5 μm particles (20024AST)

mobile phase

[A] acetonitrile: [B] acetic acid: [C] triethylamine; (100:0.3:0.25, A:B:C)

flow rate

1 mL/min

column temp.

23 °C

detector

UV, 254 nm

injection

5 μL

描述

法律資訊

Astec is a registered trademark of Merck KGaA, Darmstadt, Germany
null