跳轉至內容
Merck

HPLC Analysis of Furans on Ascentis® Express Phenyl-Hexyl

HPLC Analysis of Furans on Ascentis® Express Phenyl-Hexyl application for HPLC

材料

分析柱

產品號碼
描述
訂價

Ascentis® Express Phenyl-Hexyl,2.7μm HPLC色谱柱

2.7 μm particle size, L × I.D. 5 cm × 3 mm

標準

產品號碼
描述
訂價

5-羟甲基-2-糠醛

99%

糠醇

analytical standard

CONDITIONS

column

Ascentis Express Phenyl-Hexyl, 5 cm x 3.0 mm I.D., 2.7 μm particles (53342-U)

flow rate

0.4 mL/min

column temp.

35 °C

detector

UV, 230 nm

injection

1 μL

sample

20 μg/mL in 80:20, water: acetonitrile

mobile phase

[A] water; [B] acetonitrile; (80:20, A:B)

描述

分析報告

This application demonstrates the suitability of Ascentis Express Hexyl-Phenyl for the efficient separation of furans as listed in ASTM method D5837-99.

法律資訊

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany