跳轉至內容
Merck

GC Analysis of Agricultural Pesticides (Standard) on SLB®-5ms

GC Analysis of Agricultural Pesticides (Standard) on SLB®-5ms suitable for GC

材料

分析柱

產品號碼
描述
訂價

SLB ®-5 ms 毛细管 GC 色谱柱

L × I.D. 30 m × 0.25 mm, df 0.25 μm

CONDITIONS

column

SLB-5ms, 30 m x 0.25 mm I.D, 0.25 μm (28471-U)

oven

100 °C (1 min), 10 °C/min to 300 °C (5 min)

inj. temp.

250 °C

MSD interface

300 °C

scan range

selected ion monitoring (SIM), 7 monitoring groups used

carrier gas

helium, 1 mL/min constant

injection

1.0 μL pulsed (20 psi until 0.20 min), splitless (1.0 min)

liner

4 mm I.D., single taper

sample

29-component pesticide mix plus 1 internal standard, each at 100 ppb in toluene

描述

分析報告

This application illustrates the analysis of 29 different agricultural pesticides of different classifications and functionalities. The analysis was done by GC-MS using selected ion monitoring (SIM) to allow detection of all compounds at 100 μg/L. The
standard solution was made in toluene to increase it stability and enhance response of some of the pesticides. The extra peaks in the first portion of the run are due to interferences from the solvent.

法律資訊

SLB is a registered trademark of Merck KGaA, Darmstadt, Germany