技術
HPLC: suitable
測試參數
SPE tube/cartridge: Discovery DSC-Si, 1 gm/6 mL (52656-U)
condition: 6 mL hexane
sample addition: 0.12 g oil in 0.5 mL hexane
elution: 10 mL hexane-diethyl ether 87-13 (v/v)
eluate post-treatment: evaporation to dryness, reconstituted into 2 mL acetone
column: Discovery HS C18, 25 cm x 4.6 mm I.D., 5 μm particles (568523-U)
mobile phase: propionitrile
flow rate: 1.5 mL/min
pressure: 1117 psi (77 bar)
column temp.: 30 °C
detector: ELSD
injection: 5 μL
適合性
application for HPLC
應用
food and beverages
分析報告
Olive oil was pre-cleaned using Silica SPE, analyzed as 5% sample in acetone (w/v).
法律資訊
Discovery is a registered trademark of Merck KGaA, Darmstadt, Germany
分析物
描述
LLL
OOO
我們的科學家團隊在所有研究領域都有豐富的經驗,包括生命科學、材料科學、化學合成、色譜、分析等.
聯絡技術服務