技術
HPLC: suitable
測試參數
column: Ascentis Express C18, 10 cm x 4.6 mm I.D., 2.7 μm particles (53827-U)
mobile phase: [A] water; [B] methanol; (25:75, A:B)
flow rate: 0.8 mL/min
pressure: 2450 psi (169 bar)
column temp.: 35 °C
detector: UV, 240 nm
injection: 5 μL
sample: (1R)-2,2′-Bis[(S )-(4-methylphenyl)sulfonyl]-1,1′-binapthalene, 25 μg/mL in methanol; (1S)-2,2′-Bis[(S)-(4-methylphenyl)sulfonyl]-1,1′-binapthalene, 50 μg/mL in methanol
適合性
application for HPLC
應用
chemicals and industrial polymers
分析報告
This application demonstrates the suitability of the Ascentis Express C18 for the separation of the diastereomers of by HPLC in reversed phase mode.
法律資訊
Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany
分析物
描述
(1R)-2,2′-Bis[(S)-(4-methylphenyl)sulfinyl]-1,1′-binaphthalene, 97%
(1S)-2,2′-Bis[(S)-(4-methylphenyl)sulfinyl]-1,1′-binaphthalene, ≥96.5% (HPLC)
我們的科學家團隊在所有研究領域都有豐富的經驗,包括生命科學、材料科學、化學合成、色譜、分析等.
聯絡技術服務